Spectral emission profile and wavelength tolerances affect pulse oximeter performance
Maximilian F. ReiseriD, Andreas BreidenasseliD, Oliver AmftiD
Abstract
We investigate the effects of skin pigmentation and light source characteristics on the performance of reflective Pulse oximetry (PO) devices used in healthcare and well-being applications. We use Monte Carlo (MC) simulations to compare ideal monochromatic and realistic LED spectral emission profiles and tolerance-related wavelength shifts. The simulation covers photon transport in skin models with melanin concentrations (2.55% to 30.5%) and arterial oxygen saturations SaO2 (70% to 100%.) Accuracy was assessed by SpO2 error, root-mean-square error RMSE (Arms), and percentile tail-errors (P90, P95, and P99). Monochromatic spectral emission yielded the lowest SpO2 error (RMSE = 1.32), while LED spectral emission profiles increased errors (RMSE = 2.10). Infrared wavelength tolerances increased SpO2 RMSE by 1.1 {+/-} 0.3. SpO2 error increased with melanin concentration, from underestimation (-1.8 {+/-} 0.1%) at 2.55% melanin concentration to overestimation (+3.9 {+/-} 1.2%) at 30.5% for low SaO2 (70%) and LED spectral emission profiles. At 30.5% melanin concentration, P95 and P99 exceeded FDA and DIN EN ISO 80601-2-61 thresholds, in particular at low SaO2 (70%). Clipping SpO2 estimates at 100% resulted in an apparent RMSE decrease of up to 3%, reflecting error masking rather than real error reduction. In conclusion, LED spectral emission profiles and wavelength tolerances can amplify melanin-related bias in SpO2 estimates. Monochromatic emission and tighter wavelength control can reduce SpO2 error and should be considered in device design and regulation. Regulatory standards should discourage clipping SpO2 estimates at 100% and mandate additional metrics as RMSE fails to reflect clinically critical percentile error thresholds, i.e. P95 and P99.
02 /Specification & Classification
Click any tag to browse matching papersReproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- The effects of skin tone on photoacoustic imaging and oximetry
T. R. Else, L. Hacker, … S. Bohndiek·bioRxiv·Aug 17, 2023
- Robustness of time domain near infrared spectroscopy to variations in skin pigmentation
Michele Lacerenza, Caterina Amendola, … Alessandro M. Torricelli·Biomedical Optics ExpressPeer-reviewed·Dec 16, 2024
- Adaptive spectroscopic visible-light optical coherence tomography for human retinal oximetry
I. Rubinoff, R. V. Kuranov, … H. F. Zhang·bioRxiv·May 28, 2021
- Improved Spectral Inversion of Blood Oxygenation due to Reduced Tissue Scattering: Towards NIR-II Photoacoustic Imaging
Vinoin Devpaul Vincely, Carolyn L. Bayer·bioRxiv·Aug 8, 2024
- Optical Limits in Skin Reflectance Measurement: Quantifying Melanin-Dependent Constraints on Erythema Detection
Yemko Pryor, J. He, … Tina Lasisi·Journal of Investigative DermatologyPeer-reviewed·May 4, 2026
- Machine learning nominal max oxygen consumption from wearable reflective pulse oximetry with density functional theory
S. A. F. Al Dajani, Frédéric Laquai·bioRxiv·Jul 15, 2024