ZOnally magnified Oblique Multi-slice (ZOOM) for cardiac diffusion weighted imaging in vivo
Lars MuelleriD, Sam CoveneyiD, Isaac WatsoniD, André DöringiD, Irvin TehiD, May LwiniD, Erica Dall’ArmellinaiD, Filip SzczepankiewicziD, Derek Kenton JonesiD, Maryam AfzaliiD, Jürgen E. SchneideriD
Abstract
PurposeCardiac diffusion weighted imaging (cDWI) commonly utilizes 2D-selective RF-pulses (such as ZOOMit) to avoid aliasing artifacts when reducing the field of view (FOV). These RF-pulses, which may require parallel transmit capability, typically take several 10s of milliseconds, prolonging the echo time (TE) of the imaging sequence. Conversely, slice selection using ZOnally-magnified Oblique Multi-slice (ZOOM) (i.e. tilting the excitation relative to the refocusing RF-pulse) is an alternative for reducing FOV while shortening TE. We hypothesized that ZOOM, with appropriate choices for the RF-pulse tilt angle and thickness, can be reliably used as an alternative to ZOOMit for cardiac Diffusion Tensor Imaging (cDTI). MethodsWe scanned phantoms using full FOV and ZOOM with various parameters chosen from a 2-dimensional grid of angles and thicknesses. We identified a subset of 5 ZOOM parameters for in vivo cDTI experiments. We performed cDTI on healthy volunteers on a Siemens Prisma (6 subjects) and a Connectom (6 subjects) MR system using the 5 identified ZOOM settings, as well as ZOOMit (Prisma only) and no ZOOM (Connectom only). We compared mean diffusivity (MD), fractional anisotropy (FA), and secondary eigenvector angle (|E2A|) between scans. ResultsAverage MD, and standard deviation of MD and FA, were significantly reduced by ZOOM compared to no ZOOM on Connectom. We found no significant differences in any averaged diffusion measures between ZOOMit and ZOOM on Prisma. ConclusionCombining ZOOM with cardiac diffusion weighted imaging allows for TE to be considerably shortened (from 79ms to 67ms) while reducing FOV to avoid aliasing artifacts.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Towards genuine three-dimensional diffusion imaging with physiological motion compensation
Yishi Wang, Dehe Weng, … Zhe Zhang·bioRxiv·Sep 8, 2024
- Cardiac diffusion kurtosis imaging in the human heart in vivo using 300mT/m gradients
Maryam Afzali, Sam Coveney, … Jürgen E. Schneider·Magnetic Resonance in MedicinePeer-reviewed·Jul 3, 2025
- Evaluation of Diffusion Tensor Imaging in the Corpus Callosum on a Portable 100 mT MRI System
Philip Kenneth Lee, Suen Chen, … Zhiyong Zhang·bioRxiv·Apr 28, 2026
- Robust constrained weighted least squares for in vivo human cardiac diffusion kurtosis imaging
Sam Coveney, Maryam Afzali, … Jürgen E. Schneider·Magnetic Resonance in MedicinePeer-reviewed·Aug 24, 2025
- Optimization of a free water elimination two-compartment model for diffusion tensor imaging.
Rafael Neto Henriques, Ariel Rokem, … Marta Correia·bioRxiv·Feb 15, 2017
- Mapping Mesoscale Cardiac Architecture with Axially Swept Dithered Light-Sheet Microscopy
Milad Almasian, Alireza Saberigarakani, … Yichen Ding·bioRxiv·Oct 13, 2025