NITINOL MATERIAL PROPERTIES OF 11 COMMERCIAL PERIPHERAL STENTS DETERMINED USING INVERSE COMPUTATIONAL ANALYSIS
Eric D. AnttilaiD, Kaspars MaleckisiD, Majid JadidiiD, Anastasia S. DesyatovaiD, Jason MacTaggartiD, Alexey V. KamenskiyiD
Abstract
Stent-artery interactions are influenced by the mechanical properties of self-expanding Nitinol stents, but data on these characteristics remain limited. Eleven stents (Absolute Pro, S.M.A.R.T. Control, Misago, Zilver, Complete SE, EverFlex, Innova, Pulsar-18, LifeStent, S.M.A.R.T. Flex, and Supera) used to treat peripheral arterial disease (PAD) were subjected to axial tension, compression, three-point bending, and torsion tests, and the data on reaction forces and moments were compared with finite element simulations of the same experiments. Inverse computational analysis was used to determine austenite and martensite elasticity, transformation stretch, stresses at the start and end of transformation loading, and the start of transformation stress in compression. Uniaxial tensile tests were done on isolated struts from Absolute Pro and Zilver stents to verify the results of the inverse analysis. Our study demonstrate that Nitinol material properties are significantly different across devices. Austenite elasticity ranged 7.5-85 GPa, martensite elasticity 10-47.8 GPa, transformation stretch 1.03-1.08, the start of transformation loading stress 386-465 MPa, the end of transformation loading stress 411-535 MPa, and the start of transformation stress in compression 150-900 MPa. Nitinol of S.M.A.R.T. Control and S.M.A.R.T. Flex devices had the softest response, while Pulsar-18 had the hardest. The presented Nitinol mechanical properties of commonly used PAD stents can improve the fidelity of computational models investigating stent-artery interactions and may help improve clinical outcomes of endovascular PAD repairs through better device design.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 1
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Biomechanical Compatibility With Physiological Limb-Flexion Deformation Reduces Inflammation and Neointimal Hyperplasia After Peripheral Stenting
Alexey V. Kamenskiy, Barbara Batista de Oliveira, … Jason MacTaggart·bioRxiv·Oct 6, 2026
- Multi-objective optimisation of material properties and strut geometry for poly(L-lactic acid) coronary stents using response surface methodology
- Hemodynamic-Structural Multi-Fidelity Assessment of Endovascular Flow Diversion Strategies for Cerebral Aneurysm Treatment: An Integrated Computational and Experimental Framework
Dipu Roy, Farah Sadia Antara, … Jafrin Sultana·bioRxiv·Oct 19, 2026
- Can we obtain in vivo transmural mean hoop stress of the aortic wall without knowing patient-specific material properties and residual deformations?
- Biomechanical characterisation of thoracic ascending aorta with preserved pre-stresses
S. Parikh, K. M. Moerman, … W. Huberts·bioRxiv·Dec 2, 2022
- Cell-cultured PDMS vascular model to allow placement of implant devices
Taku Okuno, Kazuyo Ito, … Daisuke Yoshino·BMC MethodsPeer-reviewed·May 11, 2026