MR-ARFI and Beyond: Characterizing the Temporal Dynamics of Acoustic Radiation Force in Human Brain
Abstract
Purpose: Magnetic Resonance Acoustic Radiation Force Imaging (MR-ARFI) encodes focused ultrasound (FUS)-induced tissue displacements to MR phase shifts. Because these phase shifts are small, optimizing displacement sensitivity and minimizing motion artifacts requires characterization of the temporal dynamics of acoustic radiation force (ARF). Beyond sequence design, such characterization enables MR-ARFI to serve as a probe of brain viscoelastic behavior and can inform the design of therapeutic ultrasound pulsing schemes by quantifying how rapidly tissue deforms and relaxes. Methods: We developed a dynamic model assuming a gamma-variate impulse response for ARF(t), with time constant ({tau}) to be estimated. A single-shot spin echo ARFI acquisition with a bipolar motion encoding gradient (MEG) was implemented, incorporating retrospective corrections for head motion and physiological noise. A 500 kHz FUS pulse was applied in on/off blocks while systematically varying the timing between the MEG and sonication. The resulting phase time series were fit to the model to estimate time constant in five human participants. To assess sensitivity to tissue stiffness, {tau} was additionally measured in two customized gelatin phantoms of differing stiffness, characterized independently by MR elastography (MRE). Results: The average measured time constant in human participants was {tau} 3.7 {+/-} 0.27 ms. Comparing the two phantoms, the stiffer phantom exhibited a shorter {tau} (3.0-3.5 ms) than the softer phantom (4.0-4.5 ms), which tracked the MRE-derived Kelvin-Voigt viscoelastic time constants. In addition, the peak ARFI displacement was lower in the stiffer phantom (0.17 vs 0.74 m for the less stiff phantom), and a broader displacement focal spot size (FWHM 22 mm vs 15 mm). Conclusion: This time series approach with motion and physiological noise corrections validates an MR-based technique capable of characterizing the brain's temporal response to sonication-induced mechanical forces, providing a foundation for optimizing the design of ultrasound pulsing schemes for ARFI, as well as for future studies of brain biomechanics.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 0
- Organism
- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Optimization of MR-ARFI for Human Transcranial Focused Ultrasound
Morteza Mohammadjavadi, Ryan Thomas Ash, … Kim Butts·Magnetic Resonance in MedicinePeer-reviewed·May 6, 2025
- Optimal MR-ARFI encoding parameters for a range of contemporary magnetic field gradient systems
Shota Hodono, David Gordon Norris·bioRxiv·Oct 19, 2026
- VESTA: Machine Learning-Enabled Estimation of ViscoElastic Ratios from On-Axis Spatio-Temporal ARFI Features
Simina Mannan Trisha, Md Ashiqur Rahman, … Md Murad Hossain·bioRxiv·Jul 7, 2026
- Focused acoustic-radiation-force microscopy for live-cell selective nucleus deformation and nucleus viscoelasticity measurement
Natsumi Fujiwara, Hiroki Okita, … Hirotsugu Ogi·bioRxiv·Oct 23, 2025
- Validating time-resolved fMRI with sub-second event durations
Alvin P.H. Wong, Esther Wu, … Christopher Lee Asplund·bioRxiv·Jul 30, 2023
- Development of an MR-guided focused ultrasound (MRgFUS) lesioning approach for small and deep structures in the rat brain
C. Cornelssen, A. Payne, … J. D. Rolston·bioRxiv·Oct 11, 2023