Comparison of simple RNA extraction methods for molecular diagnosis of hepatitis C virus in plasma

Sayamon HongjaiseeiD, Yosita Jabjainai, Suthasinee Sakset, Kanya PreechasuthiD, Nicole Ngo‐Giang‐HuongiD, Woottichai KhamduangiD

PDFDOIPreprint

Abstract

We evaluated the efficacy of four simple RNA extraction methods for the detection of hepatitis C virus (HCV) in plasma samples: silica-membrane-based, magnetic beads-based, boiling with Diethyl Pyrocarbonate (DEPC)-treated distilled water or a commercial lysis buffer. HCV RNA was detected using both real time reverse transcription polymerase chain reaction (RT-PCR) and reverse transcription loop mediated isothermal amplification (RT-LAMP). The magnetic beads-based extraction can be used as an alternative RNA extraction method for on-site HCV detection. Boiling with DEPC-treated distilled water was not appropriate for low HCV load samples and boiling with a lysis buffer was not recommended.