Biomedical Optics ExpressBiophysicsPeer-reviewed journal article
Automated measurement of cardiomyocyte monolayer contraction using the Exeter Multiscope
Sharika MohananiD, David W. HorselliD, Taylor Watters, Mohammadreza GhasemiiD, Lewis Henderson, Marie Caroline MüllenbroichiD, Gil BubiD, Francis L. BurtoniD, Godfrey L. SmithiD, Alexander D. CorbettiD
Abstract
We apply a novel microscope architecture, the Exeter Multiscope, to the problem of acquiring image data in rapid succession from nine wells of a 96 well plate. We demonstrate that the new microscope can detect contraction in cardiomyocyte monolayers which have been plated into these wells. Each well is sampled using 500 x 500 pixels across a 1.4 x 1.4 mm field of view, acquired in three colours at 3.7 Hz per well. The use of multiple illumination wavelengths provides post-hoc focus selection, further increasing the level of automation. The performance of the Exeter Multiscope is benchmarked against industry standard methods using a commercial microscope with a motorised stage and demonstrates that the Multiscope can acquire data almost 40 times faster. The data from both Multiscope and the commercial systems are processed by a pixel variance algorithm that uses information from the pixel value variability over time to determine the timing and amplitude of tissue contraction. This algorithm is also benchmarked against an existing algorithm that employs an absolute difference measure of tissue contraction.
- assay: imaging
- research focus: cell type annotation
- research focus: clustering
- research focus: differential expression
- research focus: cell-cell communication
- approach: experimental assay
- approach: microscopy & imaging analysis
- Citations
- 3
- FWCI
- 1.02
- Field percentile
- top 25%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Versatile open software to quantify cardiomyocyte and cardiac muscle contraction in vitro and in vivo
L. Sala, B. J. van Meer, … F. L. Burton·bioRxiv·Jul 8, 2017
- Non-invasive quantification of contractile dynamics in cardiac cells, spheroids and organs-on-a-chip using high frequency ultrasound
E. M. Strohm, N. I. Callaghan, … C. A. Simmons·bioRxiv·Jun 28, 2022
- Fluorescence-lifetime optical electrophysiology in contracting cardiomyocytes
Euan Millar, Eline Huethorst, … Daniele Faccio·Proceedings of the National Academy of SciencesPeer-reviewed·Jun 3, 2026
- Compressive axial-integrated planar scanning (CAPS) microscopy for high-speed volumetric imaging of cardiac dynamics
Xinyuan Zhang, Jichen Chai, … Yichen Ding·bioRxiv·Apr 24, 2026
- Monitoring contractility in single cardiomyocytes and whole hearts with bio-integrated microlasers
M. Schubert, L. Woolfson, … M. C. Gather·bioRxiv·Jun 17, 2019
- In vivo multiphoton microscopy of cardiomyocyte calcium dynamics in the beating mouse heart
D. M. Small, J. S. Jones, N. Nishimura·bioRxiv·Jan 22, 2018