AutoSpill: A method for calculating spillover coefficients to compensate or unmix high-parameter flow cytometry data
Carlos P. RocaiD, Oliver T. BurtoniD, Václav GergelitsiD, Teresa Prezzemolo, Carly E. WhyteiD, Richard HalpertiD, Łukasz KreftiD, James H. CollieriD, Alexander BotzkiiD, Josef ŠpidleniD, Stéphanie Humblet‐BaroniD, Adrian ListoniD
Abstract
Compensating in classical flow cytometry or unmixing in spectral systems is an unavoidable challenge in the data analysis of fluorescence-based flow cytometry. In both cases, spillover coefficients are estimated for each fluorophore using single-color controls. This approach has remained essentially unchanged since its inception, and is increasingly limited in its ability to deal with high-parameter flow cytometry. Here, we present AutoSpill, a novel approach for calculating spillover coefficients or spectral signatures of fluorophores. The approach combines automated gating of cells, calculation of an initial spillover matrix based on robust linear regression, and iterative refinement to reduce error. Moreover, autofluorescence can be compensated out, by processing it as an endogenous dye in an unstained control. AutoSpill uses single-color controls and is compatible with common flow cytometry software, but it differs in two key aspects from current methods: (1) it is much less demanding in the preparation of controls, as it does not require the presence of well-defined positive and negative populations, and (2) it does not require manual tuning of the spillover matrix, as the algorithm iteratively computes the tuning, producing an optimal compensation matrix. Another algorithm, AutoSpread, complements this approach, providing a robust estimate of the Spillover Spreading Matrix (SSM), while avoiding the need for well-defined positive and negative populations. Together, AutoSpill and AutoSpread provide a superior solution to the problem of fluorophore spillover, allowing simpler and more robust workflows in high-parameter flow cytometry.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 56
- FWCI
- 3.85
- Field percentile
- top 5%
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- AutoSpectral improves spectral flow cytometry accuracy through optimised spectral unmixing and autofluorescence-matching at the cellular level
Oliver T. Burton, Leoni Bücken, … Adrian Liston·bioRxiv·Oct 27, 2025
- A gain and dynamic range independent index to quantify spillover spread to aid panel design in flow cytometry
D. Bhowmick, F. V. Diepen, … M. Ratliff·bioRxiv·Oct 31, 2020
- Ab initio Spillover Compensation in CyTOF Data
Qi Miao, Fang Wang, … Ken Chen·Cytometry Part APeer-reviewed·Dec 20, 2020
- spillR: Spillover Compensation in Mass Cytometry Data
Marco Guazzini, Alexander G. Reisach, … Christof Seiler·BioinformaticsPeer-reviewed·Jun 1, 2024
- Unbiased method for spectral analysis of cells with great diversity of autofluorescence spectra
J. E. G. Roet, A. M. Mikula, … R. E. Mebius·bioRxiv·Jul 28, 2023
- Unmixing Spread Estimation Based on Residual Model in Spectral Flow Cytometry
Xiangming Cai, Sara García‐García, … Juan Jesús García-Vallejo·bioRxiv·Jan 28, 2026