bioRxivNeurosciencePreprint
A 128-channel receive array with enhanced SNR performance for 10.5 tesla brain imaging
Russell LagoreiD, Alireza Sadeghi‐TarakamehiD, Andrea GrantiD, Matt WaksiD, Edward J. AuerbachiD, Steve JungstiD, Lance J. DelaBarreiD, Steen MoelleriD, Yiğitcan EryamaniD, Riccardo LattanziiD, Ilias I. GiannakopoulosiD, Luca VizioliiD, Essa S. YacoubiD, Simon SchmidtiD, Gregory J. MetzgeriD, Xiaoping WuiD, Gregor AdrianyiD, Kǎmil UǧurbiliD
Abstract
PurposeTo develop and characterize the performance of a 128-channel head array for brain imaging at 10.5 tesla and evaluate the potential of brain imaging at this unique, >10 tesla magnetic field. MethodsThe coil is composed of a 16-channel self-decoupled loop transmit/receive array with a 112-loop receive-only (Rx) insert. Interactions between the outer transmitter and the inner 112Rx insert were mitigated using coaxial cable traps placed every 1/16 of a wavelength on each feed cable, locating most preamplifier boards outside the transmitter field and miniaturizing those placed directly on individual coils. ResultsThe 128-channel array described herein achieved 77% of ultimate intrinsic SNR in the center of the brain. Transmit field maps obtained experimentally on a phantom with and without the receive array were similar and matched EM simulations, leading to FDA approval for human imaging. Anatomical and functional data, including with power demanding sequences, were acquired successfully on human volunteers. ConclusionsCounterintuitive to expectations based on magnetic fields [≤]7T, the higher channel counts provided SNR gains centrally, capturing [~]80% uiSNR. Fraction of uiSNR achieved centrally in 64Rx, 80Rx, and 128Rx arrays suggested that a plateau was being reached at 80%. At this plateau, linear to approximately quadratic B0 dependent SNR gains for the periphery and the center, respectively, were observed for 10.5T relative 7T.
- Citations
- 2
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- RF coil design strategies for improving SNR at the ultrahigh magnetic field of 10.5 Tesla
Matt Waks, Russell Lagore, … Kǎmil Uǧurbil·bioRxiv·May 23, 2024
- An 80-channel receive array for 10.5T neuroimaging: Key considerations for SNR optimization
Matt Waks, Alexander Bratch, … Alireza Sadeghi‐Tarakameh·bioRxiv·Jul 16, 2026
- An 8 Dipole Transceive and 24 Loop Receive Array for Non-Human Primate Head Imaging at 10.5T
R. L. Lagore, S. Moeller, … G. Adriany·bioRxiv·Aug 27, 2020
- A 48-Channel Receive Array Coil for Mesoscopic Diffusion-Weighted MRI of Human ex vivo Brain Imaging on the 3T Connectome Scanner
- Single Channel Strongly-Coupled Geometry Surface Coil for Small Mammal Whole Brain fMRI
Kyle Johnson, Hanbing Lu, Jason W. Sidabras·bioRxiv·Aug 7, 2026
- SNR Enhancement Considerations for Loop Receive Coils at Ultra-High Fields
Russell Lagore, Matt Waks, … Alireza Sadeghi‐Tarakameh·bioRxiv·Jun 19, 2026