Robust 3D Bloch-Siegert based B1+ mapping using Multi-Echo General Linear Modelling
Nadège CorbiniD, Julio Acosta‐CabroneroiD, Shaihan MalikiD, Martina F. CallaghaniD
Abstract
PurposeQuantitative MRI applications, such as mapping the T1 time of tissue, puts high demands on the accuracy and precision of transmit field (B1+) estimation. A candidate approach to satisfy these requirements exploits the difference in phase induced by the Bloch-Siegert Shift (BSS) of two acquisitions with opposite off-resonance frequency RF pulses. Interleaving these RF pulses ensures robustness to motion and scanner drifts, however, here we demonstrate that doing so also introduces a bias in the B1+ estimates. MethodsWe show via simulation and experiments that the amplitude of the error depends on MR pulse sequence parameters, such as TR and RF spoiling increment, but more problematically, on the intrinsic properties, T1 and T2, of the investigated tissue. To solve these problems, we present a new approach to BSS-based B1+ estimation that uses a multi-echo acquisition and a general linear model (GLM) to estimate the correct BSS-induced phase. ResultsIn line with simulations, phantom and in-vivo experiments confirmed that the GLM-based method removed the dependency on tissue properties and pulse sequence settings. It also showed greater robustness to hardware imperfections. ConclusionThe GLM-based method is recommended as a more accurate approach to BSS-based B1+ mapping.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 14
- FWCI
- 1.32
- Field percentile
- top 21%
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A 3T investigation of B1+-inhomogeneity tolerance in MP2RAGE-based R1-mapping by calculating second contrast that permits previously problematic sequence parameters
Lenno Ruijters, Torben E. Lund, Mads Sloth Vinding·Magnetic Resonance in MedicinePeer-reviewed·Nov 11, 2024
- Multi-echo Quantitative Susceptibility Mapping: How to Combine Echoes for Accuracy and Precision at 3 T
Emma Biondetti, Anita Karsa, … Karin Shmueli·Magnetic Resonance in MedicinePeer-reviewed·Jun 29, 2022
- Magnetic Resonance Fingerprinting with Combined Gradient- and Spin-echo Echo-planar Imaging: Simultaneous Estimation of T1, T2 and T2* with integrated-B1 Correction
Mahdi Khajehim, Thomas Christen, J. Jean Chen·bioRxiv·Apr 10, 2019
- How forgiving are M/EEG inverse solutions to noise level misspecification? An excursion into the BSI-Zoo.
Anuja Negi, Stefan Haufe, … Ali B. Hashemi·IEEE Transactions on Biomedical EngineeringPeer-reviewed·Jan 1, 2026
- Scout-based Multi-Echo NAvigator (SMENA) for high temporal resolution motion and B0 estimation and correction: applications to multi-echo GRE and EPTI
Nan Wang, Yi‐Cheng Lin, … Kawin Setsompop·bioRxiv·Jun 15, 2026
- An Adaptive H-Refinement Method for the Boundary Element Fast Multipole Method for Quasi-static Electromagnetic Modeling
William A. Wartman, Konstantin Weise, … Sergey N Makaroff·Physics in Medicine and BiologyPeer-reviewed·Feb 5, 2024