Chemical shift and relaxation regularisation improve the accuracy of 1H MR spectroscopy analysis
Abstract
PurposeAccurate analysis of metabolite levels from 1H MRS data is a significant challenge, typically requiring the estimation of approximately 100 parameters from a single spectrum. Signal overlap, spectral noise and common artefacts further complicate analysis, leading to instability and reports of poor agreement between different analysis approaches. One inconsistently used method to improve analysis stability is known as regularisation, where poorly determined parameters are partially constrained to take a predefined value. In this study we examine how regularisation of frequency and linewidth parameters influences analysis accuracy. MethodsThe accuracy of three MRS analysis methods was compaired: 1) ABfit, 2) ABfit-reg and 3) LCModel, where ABfit-reg is a modified version of ABfit incorporating regularisation. Accuracy was assessed on synthetic MRS data generated with random variability in the frequency shift and linewidth parameters applied to each basis signal. Spectra (N=1000) were generated across a range of SNR values (10, 30, 60, 100) to evaluate the impact of variable data quality. ResultsComparison between ABfit and ABfit-reg demonstrates a statistically significant (p 10 (p < 0.0005). Furthermore, Bland-Altman analysis shows that incorporating regularisation into ABfit enhances its agreement with LCModel. ConclusionRegularisation is beneficial for MRS fitting and accurate characterisation of the frequency and linewidth variability in vivo may yield further improvements.
- assay: multiome
- research focus: benchmarking
- research focus: variant calling
- research focus: differential expression
- research focus: gene regulatory network inference
- approach: statistical model
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Adaptive Baseline Fitting for 1H MR Spectroscopy Analysis
M. Wilson·bioRxiv·Feb 17, 2020
- Test-retest reproducibility of human brain multi-slice 1H FID-MRSI data at 9.4 T after optimization of lipid regularization, macromolecular model and spline baseline stiffness
Theresia Ziegs, Andrew Martin Wright, A Henning·bioRxiv·Feb 2, 2022
- Model-based frequency-and-phase correction of 1H MRS data with 2D linear-combination modeling.
Dunja Simičić, Helge J. Zöllner, … Georg Oeltzschner·bioRxiv·Mar 26, 2024
- A data-driven algorithm to determine 1H-MRS basis set composition
Christopher W. Davies‐Jenkins, Helge J. Zöllner, … Georg Oeltzschner·Magnetic Resonance in MedicinePeer-reviewed·Aug 16, 2025
- Towards harmonized spectral quantification in MRSI: Comparative analysis of Backward-Linear-Predicted and original 1H-FID-MRSI dephased data
Alessio Siviglia, Brayan Alves, … Bernard Lanz·Magnetic Resonance Materials in Physics Biology and MedicinePeer-reviewed·May 14, 2026
- FSL-MRS: An end-to-end spectroscopy analysis package
William T. Clarke, Charlotte Jane Stagg, Saâd Jbabdi·Magnetic Resonance in MedicinePeer-reviewed·Dec 6, 2020