bioRxivBioengineering
High-throughput optical sensing of peri-cellular oxygen in cardiac cells: system characterization, calibration, and testing
W. Li, D. McLeod, J. T. Ketzenberger, G. Kowalik, R. Russo, Z. Li, M. W. Kay, E. Entcheva
Abstract
Human-induced pluripotent stem cell-derived cardiomyocytes (hiPSC-CMs) represent a scalable experimental model relevant to human physiology. Oxygen consumption of hiPSC-CMs has not been studied in high-throughput (HT) format plates used in pre-clinical studies. Here, we provide comprehensive characterization and validation of a system for HT long-term optical measurements of peri-cellular oxygen in cardiac syncytia (human iPSC-CM and human cardiac fibroblasts), grown in glass-bottom 96-well plates. Laser-cut oxygen sensors having a ruthenium dye and an oxygen-insensitive reference dye were used. Ratiometric measurements (409nm excitation) reflected dynamic changes in oxygen, as validated with simultaneous Clark electrode measurements. Emission ratios (653nm vs. 510nm) were calibrated for percent oxygen using two-point calibration. Time-dependent changes in the Stern-Volmer parameter, Ksv, were observed during the initial 40 min of incubation, likely temperature-related. Effects of pH on oxygen measurements were negligible in the pH range of 4 to 8, with a small ratio reduction for pH>10. Time-dependent calibration was implemented, and light exposure time was optimized (0.6 to 0.8s) for oxygen measurements inside an incubator. Peri-cellular oxygen dropped to levels < 5% within 3 -10 hours for densely-plated hiPSC-CMs in glass-bottom 96-well plates. After the initial oxygen decrease, samples either settled to low steady-state or exhibited intermittent peri-cellular oxygen dynamics. Cardiac fibroblasts showed slower oxygen depletion and higher steady-state levels without oscillations, compared to hiPSC-CMs. Overall, the system has great utility for long-term HT monitoring of peri-cellular oxygen dynamics in vitro for tracking cellular oxygen consumption, metabolic perturbations, and characterization of the maturation of hiPSC-CMs.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Pericellular oxygen dynamics in human cardiac fibroblasts and iPSC-cardiomyocytes in high-throughput plates: insights from experiments and modeling
Weizhen Li, David G. McLeod, … Emilia G. Entcheva·Journal of Molecular and Cellular Cardiology PlusPeer-reviewed·Jun 11, 2025
- Response of human iPSC-cardiomyocytes to adrenergic drugs assessed by high-throughput pericellular oxygen measurements and computational modeling
Weizhen Li, Mohamadamin Forouzandehmehr, … Emilia G. Entcheva·Journal of Molecular and Cellular Cardiology PlusPeer-reviewed·Feb 3, 2026
- Multimodal on-axis platform for all-optical electrophysiology with near-infrared probes in human stem-cell-derived cardiomyocytes
A. Klimas, G. Ortiz, … E. Entcheva·bioRxiv·Feb 21, 2018
- Portable low-cost macroscopic mapping system for all-optical cardiac electrophysiology
Y. Heinson, J. L. Han, E. Entcheva·bioRxiv·Apr 24, 2022
- Sustained 10% Oxygen Promotes Atrial Rather Than Ventricular Specification During Human iPSC-Cardiomyocyte Differentiation
Sabrina Bech Mathiesen, Frederik Adam Bjerre, … Ditte Caroline Andersen·bioRxiv·Sep 21, 2026
- Novel 2D/3D Hybrid Organoid System for High-Throughput Drug Screening in iPSC Cardiomyocytes
Jordann Lewis, Basil Yaseen, … Anita P. Saraf·TherapeuticsPeer-reviewed·Jun 27, 2025