bioRxivBiochemistry
A low-cost and high-precision scanning electrochemical microscope built with open source tools
A. Guver, N. Fifita, P. Milas, M. Straker, M. Guy, K. Green, T. Yildirim, I. Unlu, V. Yigit, B. Ozturk
Abstract
A low-cost Scanning Electrochemical Microscope (SECM) was built with a 0.6 pA current measurement capability potentiostat and submicron resolution motorized stage, using open source software and hardware tools. The high performance potentiostat with a Python graphical user interface was built based on an open source project. Arduino boards, stepper motors, a manual XY micromanipulator stage, 3D printed couplers and gears were used in building the motorized stage. An open source motor control software was used for moving the motorized stage with high precision. An inverted microscope was utilized for viewing a standard microelectrode while scanning. The setup was tested in the formation of a map of electrochemical signals from an array of pores on a parafilm membrane. As the setup will be used in future biosensing experiments, DNA hybridization detection experiments were also performed with the setup.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Can Single Cell Respiration be Measured by Scanning Electrochemical Microscopy (SECM)?
K. Cremin, G. Meloni, … P. R. Unwin·bioRxiv·Apr 24, 2023
- An open-source, high resolution, automated fluorescence microscope
A. C. Zehrer, A. Martin-Villalba, … H. Ewers·bioRxiv·May 31, 2023
- Low-temperature integration of printed electrochemical sensors into OSTE+ micrfluidics for oxygen monitoring and sensing
Denise Marrero, Ferran Pujol‐Vila, … Xavi Illa·Lab on a ChipPeer-reviewed·Jan 1, 2026
- Super-Resolution Electrochemical Impedance Imaging with a 100 x 100 CMOS Sensor Array
K. Hu, C. E. Arcadia, J. K. Rosenstein·bioRxiv·Aug 19, 2021
- LATEER: Low-Cost Open-Source Platform for Electrical Stimulation and TEER Measurement in Human Cardiomyocytes
Florencia De Lillo, Joaquín Smucler·bioRxiv·Oct 9, 2026
- Parallel 1024-ch Cyclic Voltammetry on Monolithic CMOS Electrochemical Detector Array
K. A. White, G. Mulberry, B. N. Kim·bioRxiv·Oct 10, 2019