Musculoskeletal fat imaging and quantification by high-resolution metabolite cycling magnetic resonance spectroscopic imaging at 3 T: A fast method to generate separate distribution maps of lipid components
A. A. Alhulail, D. A. Patterson, P. Xia, X. Zhou, C. Lin, M. A. Thomas, U. Dydak, U. E. Emir
Abstract
PurposeTo provide a rapid, non-invasive fat quantification technique capable of producing separate lipid component maps.\n\nMethodsThe calf muscles in 5 healthy adolescents (age 12-16 years; BMI = 20 {+/-} 3 Kg/m2) were scanned by two different fat fraction (FF) quantification methods. A high-resolution, density-weighted concentric ring trajectory (DW-CRT) metabolite cycling (MC) magnetic resonance spectroscopic imaging (MRSI) technique was implemented to collect data with 0.25 mL resolution within 3 minutes and 16 seconds. For comparative purposes, the standard Dixon technique was performed. The two techniques were compared using structural similarity (SSIM) analysis. Additionally, the difference in the distribution of each lipid over the adolescent calf muscles was assessed based on the MRSI data.\n\nResultsThe proposed MRSI technique provided individual FF maps for eight musculoskeletal lipids identified by LCModel analysis (L09, L11, L13, L15, L21, L23, L53, and L55) with mean SSIM indices of 0.19, 0.04, 0.03, 0.50, 0.45, 0.04, 0.07, and 0.12, respectively compared to that of Dixons FF map. Further analysis of voxels with zero SSIM demonstrated an increased sensitivity of FF lipid maps from data acquired using this MRSI technique over the standard Dixon technique. The trend of lipid spatial distribution over calf muscles was consistent with previously published findings in adults.\n\nConclusionThe advantages of this MRSI technique make it a useful tool when individual lipid FF maps are desired within a short scanning time.
02 /Specification & Classification
Click any tag to browse matching papers- Organism
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Fast in vivo 23Na imaging and T2* mapping using accelerated 2D-FID magnetic resonance spectroscopic imaging at 3 T: Proof of concept and reliability study
A. A. Alhulail, P. Xia, … U. E. Emir·bioRxiv·May 25, 2020
- Lipid suppressed and tissue-fraction corrected metabolic distributions in human central brain structures using 2D 1H Magnetic Resonance Spectroscopic Imaging at 7 tesla
Alex A. Bhogal, Tommy A.A. Broeders, … Jannie P. Wijnen·Brain and BehaviorPeer-reviewed·Nov 20, 2020
- Omics Scale Quantitative Mass Spectrometry Imaging of Lipids in Brain Tissue using a Multi-Class Internal Standard Mixture
M. Vandenbosch, S. M. Mutuku, … S. R. Ellis·bioRxiv·Jun 21, 2023
- Magnetic resonance spectroscopy analysis of intramyocellular lipid composition in lipodystrophic patients and athletes
David B Savage, Laura Watson, … Alison Sleigh·bioRxiv·Aug 1, 2018
- Test-retest reproducibility of human brain multi-slice 1H FID-MRSI data at 9.4 T after optimization of lipid regularization, macromolecular model and spline baseline stiffness
Theresia Ziegs, Andrew Martin Wright, A Henning·bioRxiv·Feb 2, 2022
- Impacts of retrospective lipid suppression on metabolite quantification in preclinical proton MR spectroscopic imaging
Phan Tan Toi, Brayan Alves, … Cristina Cudalbu·bioRxiv·Sep 21, 2026