Photoacoustics-guided Real-Time Closed-loop Control of Magnetic Microrobots through Deep Learning
Richard NauberiD, Johanna Hoppe, David Castellanos RoblesiD, Mariana Medina‐SáncheziD
Abstract
Medical microrobots promise to increase the efficacy and reduce the invasiveness of certain medical procedures in the future. Real-time tracking of the microrobot, actuation, and closed-loop control of its position under in vivo conditions is crucial to fulfill the task at hand. We present a system for closed-loop control of magnetic microrobots using dual-mode ultrasound and photoacoustic imaging. It employs GPU-accelerated beamforming and tracking to achieve real-time operation with a closed-loop cycle time of 100 ms. Artifacts from simultaneous imaging and magnetic actuation are suppressed through time-multiplexing. To address the challenge of detecting microrobots in low-contrast, strong-background images, we implemented real-time Deep Learning-based tracking. A custom dataset of various types of microrobots is curated from long-duration closed-loop control measurements and employed to fine-tune a pre-trained detection model. We introduce a platform for real-time closed-loop control of microrobots and demonstrate its performance with a 300 m spiral-shaped microrobot following a figure-of-8 shape under photoacoustic imaging guidance. The localization error is evaluated against an optical reference measurement. Our results show that photoacoustic-based tracking significantly outperforms ultrasound tracking, with the deep learning approach further reducing missed detections. This demonstrates the algorithms ability to generalize to a previously unseen type of microrobot. We envision this platform to advance medical microrobotics research by providing real-time closed-loop control of untethered microrobots under deep tissue.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 17
- FWCI
- 19.33
- Field percentile
- top 1%
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Influence of hyperparameters on the performance of deep learning-based microrobotic localization under phantom tissue
Johanna Hoppe, Richard Nauber, … Mariana Medina‐Sánchez·bioRxiv·Aug 13, 2025
- Real-time Color Flow Mapping of Ultrasound Microrobots
Cornel Dillinger, Ahilan Rasaiah, … Daniel Ahmed·Science AdvancesPeer-reviewed·Jul 18, 2025
- In vivo imaging of swimming micromotors using hybrid high-frequency ultrasound and photoacoustic imaging
A. Aziz, J. Holthof, … M. Medina-Sanchez·bioRxiv·Jun 15, 2020
- Model-Based Reinforcement Learning for Ultrasound-Driven Autonomous Microrobots
Mahmoud Medany, Lorenzo Piglia, … Daniel Ahmed·Nature Machine IntelligencePeer-reviewed·Jun 26, 2025
- A Tumbling Magnetic Microrobot System for Biomedical Applications
E. E. Niedert, C. Bi, … D. J. Cappelleri·bioRxiv·Jun 4, 2020
- Real-time optoacoustic tracking of single moving micro-objects in deep tissue-mimicking phantoms
A. Aziz, M. Medina-Sanchez, … O. G. Schmidt·bioRxiv·May 16, 2019