Artificial OrgansBioengineeringPeer-reviewed journal article
Development and Validation of a Continuous Real-Time Optical Sensor for Indocyanine Green Clearance Measurement During Ex-Vivo Perfusion of Human Livers
Edmund N.J. Derwent, Charles W.G. RisbeyiD, Anita NiuiD, Paul YousifiD, Nidula Fonseka, Spencer Curry, Chloe Kwok-Long SeowiD, Isaac Ng, Geoffrey William McCaughaniD, Michael Crawford, Carlo PulitanòiD, Daniel BabekuhliD
- Citations
- 0
- FWCI
- 0.00
- Field percentile
- 15th percentile
Abstract
Liver transplantation remains the only curative treatment for end-stage liver failure, yet its impact is constrained by organ shortages and graft non-utilisation. Machine perfusion (MP) enables ex-vivo assessment of donated livers; however, existing viability criteria rely on intermittent sampling, reducing temporal resolution and accuracy. Indocyanine green (ICG), a clinically validated dye cleared exclusively by hepatocytes, provides a continuous index of hepatic function beyond initial injury. Accordingly, we present a non-invasive, clamp-on optical sensor that enables continuous, real-time quantification of ICG clearance during MP. The sensor consists of a clamp-on module with an 808nm laser and phototransistor connected to a microcontroller-based unit and computer for real-time plotting. The raw phototransistor signal was linearised to a unitless absorbance signal proportional to perfusate ICG; bi-exponential fitting yielded plasma disappearance rate (PDRbi, %/min) and the 15-minute residual fraction (R15). Across 10 whole and 3 split human livers (45 boluses; 13 paired with spectrophotometry), the sensor closely matched spectrophotometric measurements (mean R2 = 0.994; range 0.983-0.999). The sensor resolved expected physiological trends: ICG clearance increased with temperature (PDRbi: 8.2%/min (subnormothermic MP) to 22.6%/min (normothermic MP) (n=4); 9.3%/min (32{degrees}C) to 11.9%/min (36{degrees}C) (n=1)). The sensors continuous signal traces also revealed early mixing dynamics and medication-related effects that are missed by intermittent sampling. This optical sensor enables accurate, real-time monitoring of ICG clearance during ex-vivo perfusion. The ex-vivo setting is uniquely positioned to validate ICG clearance models and enhance clinical interpretation.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Real time evaluation of the liver microcirculation by whole organ machine perfusion within an MRI system
Zainab L. Rai, Natalie Holroyd, … Brian Ritchie Davidson·NMR in BiomedicinePeer-reviewed·Feb 6, 2026
- Prediction of survival after partial hepatectomy using a physiologically based pharmacokinetic model of indocyanine green liver function tests
A. Köller, J. Grzegorzewski, … M. König·bioRxiv·Jun 15, 2021
- Physiologically based modeling of the effect of physiological and anthropometric variability on indocyanine green based liver function tests
Adrian Köller, Jan Grzegorzewski, Matthias König·Frontiers in PhysiologyPeer-reviewed·Nov 22, 2021
- Improving retention time of Indocyanine Green for in vivo two-photon microscopy using liposomal encapsulation
Alankrit Tomar, Noah Stern, … Andrew K. Dunn·bioRxiv·Mar 20, 2025
- Monitoring deep-tissue oxygenation with a millimeter-scale ultrasonic implant
S. Sonmezoglu, J. R. Fineman, … M. M. Maharbiz·bioRxiv·Mar 5, 2021
- A MICROFLUIDIC DEVICE FOR LONG-TERM MAINTENANCE OF ORGANOTYPIC LIVER CULTURES
J. M. de Hoyos-Vega, H. Hong, … A. Revzin·bioRxiv·Jun 18, 2022