Inkjet-printed graphene multielectrode arrays: an accessible platform for in vitro cardiac electrophysiology
Jairo Lumpuy‐CastilloiD, Yujie FuiD, Alan Eduardo Avila RamireziD, Kateryna Solodka, Jiantong LiiD, Óscar LorenzoiD, Erica ZeglioiD, Leonardo D. GarmaiD
Abstract
In vitro models have now become a realistic alternative to animal models for cardiotoxicity assessment. However, the cost and expertise required to implement in vitro electrophysiology systems to study cardiac cells poses a strong obstacle to their widespread use. This study presents a novel, cost-effective approach for in vitro cardiac electrophysiology using fully-printed graphene-based microelectrode arrays (pGMEAs) coupled with an open-source signal acquisition system. We characterized the pGMEAs electrical properties and biocompatibility, observing low impedance values and cell viability. We demonstrated the platforms capability to record spontaneous electrophysiological activity from HL-1 cell cultures, and we monitored and quantified their responses to chemical stimulation with noradrenaline. This study demonstrates the feasibility of producing fully-printed, graphene-based devices for in vitro electrophysiology. The accessible and versatile platform we present here represents a step further in the development of alternative methods for cardiac safety screening.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 2
- FWCI
- 0.63
- Field percentile
- top 39%
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Graphene Biointerface for Cardiac Arrhythmia Diagnosis and Treatment
Z. Lin, D. Kireev, … I. R. Efimov·bioRxiv·Jun 28, 2022
- Development of human iPSC-derived cardiomyocytes-based heart-on-a-chip with customized multi-electrode array for drug-induced cardiotoxicity evaluations
Geonho Choi, Ki‐Suk Kim, … Sungho Ko·bioRxiv·Sep 16, 2025
- Generation and Culture of Cardiac Microtissues in a Microfluidic Chip with a Reversible Open Top Enables Electrical Pacing, Dynamic Drug Dosing and Endothelial Cell Co-Culture
A. Vivas, C. IJspeert, … A. D. v. d. Meer·bioRxiv·Nov 1, 2021
- Chronic cardiotoxicity assessment by cell optoporation on microelectrode arrays
G. Iachetta, G. Melle, … M. Dipalo·bioRxiv·Jun 20, 2022
- Nanocrown electrodes for reliable and robust intracellular recording of cardiomyocytes and cardiotoxicity screening
- FRESH™ 3D bioprinted cardiac tissue, a bioengineered platform for in-vitro toxicology and pharmacology