IMU-based upper limb kinematics validation: sensor fusion optimisation, sensor-to-segment alignment, and joint angle accuracy.

Arthur FabreiD, Graham Kerr, Peter PivonkaiD, Wolbert van den HoorniD

PDFDOI

Abstract

Conventional clinical assessment of upper limb kinematics relies on goniometry or visual estimation, methods that provide limited information on multi-segment 3D kinematics and can have poor inter-rater reliability. Inertial measurement units (IMUs) represent a viable alternative; however, extracting anatomically meaningful joint angles from skin-mounted sensors remains challenging owing to sensor-to-segment misalignment, soft-tissue artefact, and suboptimal orientation estimation under dynamic conditions. This study developed and validated a multi-segment IMU-based framework for upper limb kinematic assessment, addressing three stages of the measurement pipeline: (i) sensor fusion gain optimisation as a function of task mean segment angular velocity; (ii) accuracy of an anatomical landmark registration-based alignment method; and (iii) comparative evaluation of landmark-based alignment against manual and neutral-pose alignment approaches. Twenty participants with typical shoulder function performed clinical and functional upper limb tasks, recorded simultaneously with an optoelectronic motion capture system and six IMUs tracking the thorax, clavicle, scapula, humerus, forearm, and hand. Orientation estimation used a gradient-descent complementary filter with velocity-dependent gain modulation. Sensor-to-segment alignment was achieved via manual, neutral-pose, and landmark registration methods. Mean sensor orientation errors ranged from 1.5 to 3 degrees across segments/tasks and increased with angular velocity. Landmark registration produced substantially lower joint angle errors than manual and neutral-pose alignment across all joints (root mean square error range: 3.7-7.8 degrees; manual: 8.3-24.2 degrees; neutral: 15.5-35.5 degrees). These results indicate that IMU-derived kinematic accuracy is context-dependent, varying with joint and movement type, suggesting that accuracy should be established for the specific movements and joints relevant to each clinical application.

Citations
0

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISBorno '22Rapp '20Castellaz '26Diwakarredd '24Serra-Hsu '22Subash '22
77% semantic matchBioengineering
M. Al Borno, J. O'Day, V. Ibarra+7 more·Journal of NeuroEngineering and Rehabilitation·Feb 20, 2022