Eddy current-induced artifacts correction in high gradient strength diffusion MRI with dynamic field monitoring: demonstration in ex vivo human brain imaging
G. Ramos Llorden, D. Park, J. E. Kirsch, A. Scholz, B. Keil, C. Maffei, H.-H. Lee, B. Bilgic, B. Edlow, C. Mekkaoui, A. Yendiki, T. Witzel, S. Y. Huang
Abstract
PurposeTo demonstrate the advantages of spatiotemporal magnetic field monitoring to correct eddy current-induced artifacts (ghosting and geometric distortions) in high gradient strength diffusion MRI (dMRI). MethodsA dynamic field camera with 16 NMR field probes was used to characterize eddy current fields induced from diffusion gradients for different gradients strengths (up to 300 mT/m), diffusion directions, and shots in a 3D multi-shot EPI sequence on a 3T Connectom scanner. The efficacy of dynamic field monitoring-based image reconstruction was demonstrated on high-resolution whole brain ex vivo dMRI. A 3D multi-shot image reconstruction framework was informed with the actual nonlinear phase evolution measured with the dynamic field camera, thereby accounting for high-order eddy currents fields on top of the image encoding gradients in the image formation model. ResultsEddy current fields from diffusion gradients at high gradient strength in a 3T Connectom scanner are highly nonlinear in space and time, inducing high-order spatial phase modulations between odd/even echoes and shots that are not static during the readout. Superior reduction of ghosting and geometric distortion was achieved with dynamic field monitoring compared to ghosting approaches such as navigator- and structured low-rank-based methods or MUSE, followed by image-based distortion correction with eddy. Improved dMRI analysis is demonstrated with diffusion tensor imaging and high-angular resolution diffusion imaging. ConclusionStrong eddy current artifacts characteristic of high gradient strength dMRI can be well corrected with dynamic field monitoring-based image reconstruction, unlike the two-step approach consisting of ghosting correction followed by geometric distortion reduction with eddy.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- High-fidelity, high-spatial-resolution diffusion MRI of the ex-vivo whole human brain on the 3T Connectom scanner using structured low-rank EPI ghost correction
Gabriel Ramos‐Llordén, Rodrigo A. Lobos, … Susie Y. Huang·bioRxiv·Aug 1, 2021
- Diffusion MRI Head Motion Correction Methods are Highly Accurate but Impacted by Denoising and Sampling Scheme
Matthew Cieslak, Philip A. Cook, … Theodore Daniel Satterthwaite·Human Brain MappingPeer-reviewed·Jan 30, 2024
- Empirical field mapping for gradient nonlinearity correction of multi-site diffusion weighted MRI
Colin B. Hansen, Baxter P. Rogers, … Bennett A. Landman·Magnetic Resonance ImagingPeer-reviewed·Nov 19, 2020
- Slice-direction geometric distortion evaluation and correction with reversed slice-select gradient acquisitions
Anna I. Blazejewska, Thomas Witzel, … Jon̈athan R. Polimeni·NeuroImagePeer-reviewed·Oct 23, 2022
- Exploiting Four-way Phase-encoding Benefits for Robust Detection and Correction of EPI artifacts: Application to Residual Ghosts in Diffusion MRI.
Anh Thai, Lin‐Ching Chang, … Mustafa Okan Irfanoglu·Magnetic Resonance ImagingPeer-reviewed·Jul 7, 2025
- DACO: Distortion/artefact correction for diffusion MRI data in an integrated framework
Yung‐Chin Hsu, Wen‐Yih Isaac Tseng·bioRxiv·Jul 6, 2021