Performances of NeuMoDx, a random-access system for HBV-DNA and HCV-RNA quantification
J. Besombes, C. Pronier, C. Lefevre, G. Lagathu, A. Maillard, C. Grolhier, V. Thibault
Abstract
Viral loads (VL) monitoring for hepatitis B and C is essential to evaluate disease progression and treatment response. Automated, random-access rapid systems are becoming standard to provide reliable VL to clinicians. The aim of this study was to evaluate the analytical performances of the recently launched NeuMoDx for HBV-DNA and HCV-RNA quantification. Clinical samples routinely quantified on the Beckman-Veris system were either retrospectively (frozen samples; HBV n=178, HCV n=249), or in parallel (fresh primary tubes; HBV n=103, HCV n=124) tested using NeuMoDx. Linearity range was assessed on serial dilutions of high tittered plasmas containing different genotypes for HBV (A-E, n=10) and HCV (1a-b, 2-5, n=12). Overall test failure, mostly internal control amplification failure, was 2.3% and was not influenced by matrix types. For HBV-VL, Kappa agreement was 74%, with 27 (12.6%) discrepancies. Correlation between HBV assays on 72 quantified samples by both methods was excellent (r=0.963) with a mean bias (NeuMoDx-Veris) of 0.21 log IU/mL. For HCV-VL, Kappa agreement reached 94%, with 9 (2.8%) discrepancies. The r-correlation factor between assays on 104 samples was 0.960 with a mean bias of -0.14 log IU/mL (NeuMoDx-Veris). Serial dilutions confirmed the claimed linear ranges for all HBV and HCV genotypes. The mean turnaround time was 72 [55-101] for HBV and 96 [78-133] for HCV. These results obtained on the NeuMoDx confirmed the overall good functionality of the system with a short turn-around-time, full traceability and easy handling. These results on HBV- and HCV-VL look promising and should be challenged with further comparisons.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Performance of the cobas(R) HBV RNA Automated Investigational Assay for the Detection and Quantification of Circulating HBV RNA in Chronic HBV Patients
C. SCHOLTES, A. T. Hamilton, … F. Zoulim·bioRxiv·Mar 9, 2022
- Multi-site Performance Evaluation of the cobas 5800 System and Comparison to the cobas 6800/8800 Systems for Quantitative Measurement of HBV, HCV, and HIV-1 Viral Load.
J. Tschape, A. Cobernuss-Rahn, … P. Gohl·bioRxiv·Aug 9, 2022
- Systematic review of the validity of the hepatitis B virus detection tests in blood banks from 2000 to 2018.
C. P. Orrego-Marin, J. A. Cardona-Arias, … J. C. Zapata·bioRxiv·Apr 17, 2020
- CCQM-P199: Interlaboratory comparability study of HIV-1 RNA copy number quantification
Alison S. Devonshire, Eloise J. Busby, … Jim F. Huggett·bioRxiv·Apr 12, 2024
- A multiscale model of the action of a capsid assembly modulator for the treatment of chronic hepatitis B
Sarafa A. Iyaniwura, Tyler Cassidy, … Alan S. Perelson·PLoS Computational BiologyPeer-reviewed·May 6, 2025
- Comparison of simple RNA extraction methods for molecular diagnosis of hepatitis C virus in plasma
Sayamon Hongjaisee, Yosita Jabjainai, … Woottichai Khamduang·DiagnosticsPeer-reviewed·Jun 30, 2022