Multi-site Performance Evaluation of the cobas 5800 System and Comparison to the cobas 6800/8800 Systems for Quantitative Measurement of HBV, HCV, and HIV-1 Viral Load.
J. Tschape, A. Cobernuss-Rahn, S. Boyle, N. T. Parkin, B. LaBrot, S. Aslam, S. Young, P. Gohl
Abstract
BackgroundThe cobas(R) 5800 System ("cobas 5800") is a new low to mid-throughput PCR-based nucleic acid testing system which performs both qualitative and quantitative testing, including viral load (VL) determination. cobas 5800 shares numerous design elements and technical characteristics with the existing cobas 6800/8800 Systems. Study DesignWe compared HBV, HCV, and HIV-1 VL results from cobas 5800 in three different laboratories to those from the same specimens tested on a cobas 6800 system. We also assessed cobas 5800 assay reproducibility by repetitive testing of specimens with VL close to values used as thresholds for patient management or classification. ResultsThe correlation between VL measurements generated using cobas 5800 vs 6800 was extremely high, with r2 correlation coefficients between 0.990 and 0.999 for the three targets at the different sites. The slope of the Deming regression line ranged from 0.994 (HBV, site 3) to 1.025 (HIV-1, site 1). The standard deviation values ranged from 0.04 to 0.19 log10 IU/mL for HBV, 0.05 to 0.31 log10 IU/mL for HCV, and 0.05 to 0.31 log10 copies/mL for HIV-1. In general, variability was higher at lower VL. Between 98.6% and 100% of results fell within the allowable total difference zone that defines expected variability on the existing 6800/8800 system. ConclusionsThis multi-site comparison study demonstrates equivalent performance of the new cobas 5800 system compared to cobas 6800. This establishes cobas 5800 as a new option for low to mid-throughout laboratories seeking to optimize efficiency of their viral molecular testing.
02 /Specification & Classification
Click any tag to browse matching papers- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Performances of NeuMoDx, a random-access system for HBV-DNA and HCV-RNA quantification
J. Besombes, C. Pronier, … V. Thibault·bioRxiv·Oct 14, 2020
- Performance of the cobas(R) HBV RNA Automated Investigational Assay for the Detection and Quantification of Circulating HBV RNA in Chronic HBV Patients
C. SCHOLTES, A. T. Hamilton, … F. Zoulim·bioRxiv·Mar 9, 2022
- Multicenter comparison of the Cobas 6800 system with the RealStar RT-PCR kit for the detection of SARS-CoV-2
M.-L. Chaix, M. Wirden, … B. Visseaux·bioRxiv·Jun 29, 2020
- Whatman FTA(R) cards versus plasma specimens for the quantitation of HIV-1 RNA using two Real-Time PCR assays
Abdourahamane Yacouba, Malika Congo-Ouédraogo, … Lassana Sangaré·Access MicrobiologyPeer-reviewed·Jun 17, 2020
- CCQM-P199: Interlaboratory comparability study of HIV-1 RNA copy number quantification
Alison S. Devonshire, Eloise J. Busby, … Jim F. Huggett·bioRxiv·Apr 12, 2024
- Test Agreement Between Roche Cobas 6800 and Cepheid GeneXpert Xpress SARS-CoV-2 Assays at High Cycle Threshold Ranges
K. J. Broder, A. Babiker, … C. S. Kraft·bioRxiv·May 5, 2020