Trace-based correction of breathing-induced field fluctuations in T2*-weighted imaging of the spinal cord
S Johanna Vannesjo, Stuart ClareiD, Lars KasperiD, Irene TraceyiD, Karla L. MilleriD
Abstract
PurposeSpinal cord MRI at ultra-high field is hampered by time-varying magnetic fields associated with the breathing cycle, giving rise to ghosting artifacts in multi-shot acquisitions. Here, we suggest a correction approach based on linking the signal from a respiratory bellows to field changes inside the spinal cord. The information is used to correct the data at the image reconstruction level.\n\nMethodsThe correction was demonstrated in the context of multi-shot T2*-weighted imaging of the cervical spinal cord at 7T. A respiratory trace was acquired during a high-resolution multi-echo gradient-echo sequence, used for structural imaging and quantitative T2* mapping, and a multi-shot EPI time series, as would be suitable for fMRI. The coupling between the trace and the breathing-induced fields was determined by a short calibration scan in each individual. Images were reconstructed with and without trace-based correction.\n\nResultsIn the multi-echo GRE, breathing-induced fields caused severe ghosting in the long echo time images, which led to a systematic underestimation of T2* in the spinal cord. The trace-based correction reduced the ghosting and increased the estimated T2* values. Breathing-related ghosting was also observed in the multi-shot EPI images. The correction largely removed the ghosting, thereby improving the temporal signal-to-noise ratio of the time series.\n\nConclusionsTrace-based retrospective correction of breathing-induced field variations can reduce ghosting and improve quantitative metrics in multi-shot structural and functional T2*-weighted imaging of the spinal cord. The method is straightforward to implement and does not rely on sequence modifications or additional hardware beyond a respiratory bellows.
- Citations
- 1
- FWCI
- 0.17
- Field percentile
- top 44%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Impact of Autocalibration Method on Accelerated Echo-Planar Imaging of the Cervical Spinal Cord at 7T
A. C. Seifert, J. Xu·bioRxiv·Mar 8, 2022
- Integration of a radiofrequency coil and commercial field camera for ultra-high-field MRI
K. M. Gilbert, P. Dubovan, … C. A. Baron·bioRxiv·Sep 27, 2021
- Mesoscopic whole-brain T2*-weighted and associated quantitative MRI in healthy humans at 10.5 T
Jiaen Liu, Peter van Gelderen, … Xiaoping Wu·Magnetic Resonance in MedicinePeer-reviewed·Apr 7, 2026
- Advances in Spiral fMRI: A High-resolution Study with Single-shot Acquisition
Lars Kasper, Maria Engel, … Klaas P. Pruessmann·NeuroImagePeer-reviewed·Nov 17, 2021
- Magnetic Resonance Fingerprinting with Combined Gradient- and Spin-echo Echo-planar Imaging: Simultaneous Estimation of T1, T2 and T2* with integrated-B1 Correction
M. Khajehim, T. Christen, J. J. Chen·bioRxiv·Apr 10, 2019
- Feasibility of spiral fMRI based on an LTI gradient model
Nadine N. Graedel, Lars Kasper, … S. Johanna Vannesjo·NeuroImagePeer-reviewed·Oct 28, 2021