High-resolution cryo-EM reconstructions in the presence of substantial aberrations
Abstract
The beam-image shift method accelerates data acquisition in cryo-EM single particle reconstruction (cryo-EM SPR) by fast repositioning of the imaging area, but at the cost of more severe and complex optical aberrations.\n\nWe analyze here how uncorrected anti-symmetric aberrations, such as coma and trefoil, affect cryo-EM SPR results, and then infer an analytical formula quantifying information loss due to their presence that explains why Fourier-shell coefficient (FSC)-based statistics may report significantly overestimated resolution if these aberrations are not fully corrected. We validate our analysis with reference-based aberration refinement for two cryo-EM SPR datasets acquired with a 200 kV microscope in the presence of coma exceeding 40 {micro}m, and obtained 2.3 and 2.7 [A] reconstructions for 144 and 173 kDa particles, respectively.\n\nOur results provide a description of an efficient approach for assessing information loss in cryo-EM SPR data acquired in the presence of higher-order aberrations and address inconsistent guidelines regarding the level of aberrations acceptable in cryo-EM SPR experiments.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Estimation of high-order aberrations and anisotropic magnification from cryo-EM datasets in RELION-3.1
J. Zivanov, T. Nakane, S. Scheres·IUCrJPeer-reviewed·Feb 11, 2020
- High Resolution Single Particle Cryo-Electron Microscopy using Beam-Image Shift
A. Cheng, E. T. Eng, … B. Carragher·bioRxiv·Apr 26, 2018
- High-resolution cryo-EM using beam-image shift at 200 keV
J. N. Cash, S. Kearns, … M. A. Cianfrocco·IUCrJPeer-reviewed·Oct 29, 2020
- CryoEM single particle reconstruction with a complex-valued particle stack
R. Bromberg, Y. Guo, … Z. Otwinowski·bioRxiv·Jul 28, 2022
- Non-Uniformity of Projection Distributions Attenuates Resolution in Cryo-EM
P. Baldwin, D. Lyumkis·bioRxiv·May 16, 2019
- Increasing vitrification temperature improves the cryo-electron microscopy reconstruction