Evaluations of retrospective frequency and phase correction methods for single-voxel MR spectroscopy at 7T

Chu‐Yu LeeiD, Jia XuiD, Baolian Yang, Ralph NoeskeiD, Vincent A. MagnottaiD

Abstract

Subject motion and gradient heating-induced frequency and phase offsets result in spectral misalignment during single-voxel MR spectroscopy (SVS) acquisitions. Several methods have been presented to align the spectra, but their performance on a 7T system is unclear. This study aimed to evaluate the practical importance of four retrospective correction methods, namely creatine fitting, residual water, spectral registration, and cross-correlation. SVS data were collected from 127 participants (88/39 female/male, 39{+/-}15 years) using a semi-localization by adiabatic selective refocusing sequence at 7T. Changes in the spectral linewidth, signal-to-noise ratio (SNR), similarity (mean similarity matrix (SImean)), and metabolite quantification (concentration, relative Cramer-Rao lower bounds (rCRLB)) after correction were evaluated. A Wilcoxon signed rank test was used to evaluate these changes. The p-values were adjusted using the false discovery rate (q0.60). The correction only showed a weak to moderate effect of the changes in the metabolite quantification (effect size<0.60), suggesting that whether the correction results in more reliable metabolite quantification requires further validation.

PDFDOI

02 /Specification & Classification

Click any tag to browse matching papers

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISManting '26Ma '23Chhabra '26Wright '22Bacon '26Simičić '24
67% semantic matchNeuroscience
Cassia Low Manting, Atsushi Takahashi, Jyrki Ahveninen+2 more·Imaging Neuroscience·Jan 1, 2026