Validation and Comparison of a Modified CDC Assay with two Commercially Available Assays for the Detection of SARS-CoV-2 in Respiratory Specimen
Abstract
Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), the virus that causes coronavirus disease 2019 (COVID-19), has spread rapidly around the globe since it was first identified in December of 2019 in Wuhan, China. In a race to contain the infection, researchers and healthcare officials have developed several assays to help diagnose individuals with COVID-19. To help laboratories in deciding what assay to bring into testing lines, factors such as assay availability, cost, throughput, and TAT should be considered. Here we validated a modified version of the CDC assay and used it as a reference to evaluate the performance of the NeuMoDx SARS-CoV-2 and DiaSorin Simplexa Covid-19 Direct assays. In silico analysis and clinical sample testing showed that the primesr/probes designed by the CDC were specific to the SARS-CoV-2 as they accurately detected all reactive samples with an assay LoD of 200 copies/ml. The performance of the three assays were analyzed using 161 nasopharyngeal swabs specimen tested within 24 hours or 5 days from routine testing. A 100% agreement was observed between the commercial assays and the modified CDC SARS-CoV-2 assay. A deeper look at the Ct values showed no significant difference between NeuMoDx and the modified CDC SARS-CoV-2 assay, whereas DiaSorin had lower overall Ct values than the modified CDC SARS-CoV-2 assay. NeuMoDx and DiaSorin workflows were much easier to perform. NeuMoDx has the highest throughput and shortest TAT, whereas although the modified CDC SARS-CoV-2 assay has comparable throughput to DiaSorin, it has the longest hands-on time, and highest TAT.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Design of a Novel Multiplex Real Time RT-PCR Assay for SARS-CoV-2 Detection
W. Zhen, G. J. Berry·bioRxiv·Jun 4, 2020
- Comparison of the NeuMoDX, Diasorin Simplexa, Cepheid and Roche CDC SARS-CoV 2 EUA assays using nasopharyngeal/nasal swabs in universal transport media (UTM) and sputum and tracheal aspirates
R. J. Tibbetts, K. Callahan, … L. P. Samuel·bioRxiv·May 26, 2020
- Comparison of 12 molecular detection assays for SARS-CoV-2
Y. Matsumura, T. Shimizu, … M. Nagao·bioRxiv·Jun 24, 2020
- Comparative study of four SARS-CoV-2 Nucleic Acid Amplification Test (NAAT) platforms demonstrates that ID NOW performance is impaired substantially by patient and specimen type.
P. R. Lephart, M. A. Bachman, … D. W. Newton·bioRxiv·Jun 4, 2020
- Analytical and Clinical Comparison of Three Nucleic Acid Amplification Tests for SARS-CoV-2 Detection
E. Smith, W. Zhen, … G. J. Berry·bioRxiv·May 14, 2020
- Comparison of three TaqMan Real-Time Reverse Transcription-PCR assays in detecting SARS-CoV-2
Y. Xiao, Z. Li, … J. Li·Journal of Virological MethodsPeer-reviewed·Feb 1, 2021