Sensitivity of Pulmonary 4D Flow MRI Biomarkers to Controlled Spatial and Temporal Degradation in a CFD Benchmark
Aitor Zubillaga-UnsainiD, César Caballero‐GaudesiD, Jesús Ruı́z-CabelloiD
Abstract
Purpose: To quantify, in a computational fluid dynamics (CFD) benchmark, how spatial degradation, temporal acquisition and noise change the error of plane-based pulmonary-artery flow endpoints. Methods: Twenty-two historical, non-converged CFD simulations (11 geometry families) were voxelised at 2 mm with 80 frames per cycle. Simulations with numerical divergence were excluded, leaving 13 (7 families). Fields were degraded to 4 and 8 mm and with noise, plane waveforms were sampled at 20 phases with one-interval averaging, and errors were computed against the noise-free reference waveform on 66 geometric planes. Family-mean paired differences are reported with t-based 95% confidence intervals under an analysis plan fixed in advance. Results: Spatial degradation from 2 to 4 mm increased full-waveform nRMSE by 0.0032 [0.0021, 0.0043], in 7 of 7 families; cycle-mean flow bias accounted for 23% of the squared error, against 60% at 8 mm. With SNR scaled to voxel volume, the increase was not resolved (0.0002 [-0.0012, 0.0015]). Twenty-phase acquisition increased PI error (0.0223 [0.0214, 0.0233]) and diastolic-trough timing error (0.63 [0.52, 0.74] frames) more than spatial degradation, and its point estimates for these endpoints were larger with every recovery method; the nRMSE ordering and peak timing depended on the recovery method. The implementation of the downsampling operator mattered: an alternative, corner-aligned implementation of the same 4 mm degradation gave a larger spatial effect (nRMSE difference 0.0010 [0.0002, 0.0017]). Conclusion: In this benchmark, a 2 to 4 mm degradation produced a small, consistent, mostly shape-related nRMSE increase that depended on the noise model, whereas 20-phase acquisition changed PI error and trough timing more. The results describe benchmark behaviour, not clinical acquisition requirements.
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Effect of Wall Motion Sampling on CFD-Derived Left Atrial Flow Metrics
Yvonne Stöcker, Manuel Guerrero-Hurtado, … Manuel García‐Villalba·bioRxiv·Oct 24, 2025
- Biomarker-Aware Super-Resolution for 4D Flow MRI in a Synthetic Pulmonary-Artery Benchmark
Aitor Zubillaga-Unsain, Adrian Lluveras-Sires, … Jesús Ruı́z-Cabello·bioRxiv·Sep 28, 2026
- Improving the accuracy of cerebral blood flow measurement by phase contrast MRI
Xiuli Yang·bioRxiv·Aug 13, 2024
- Left atrial flow and thrombosis risk from 4D CT contrast dynamics by physics-informed neural network and indicator dilution theory
Bahetihazi Maidu, Alejandro Gonzalo, … Juan C. del Álamo·bioRxiv·Apr 3, 2026
- Computational Fontan Analysis: Preserving accuracy while expediting workflow
Xiaolong Liu, Seda Aslan, … Axel Krieger·World Journal for Pediatric and Congenital Heart SurgeryPeer-reviewed·Apr 21, 2022
- Hemodynamics in Patients with Aortic Coarctation: A Comparison of in vivo 4D-Flow MRI and FSI Simulation
P. J. Nair, M. R. Pfaller, … A. L. Marsden·bioRxiv·Feb 13, 2023