Three-Dimensional Fluid-Structure Interaction Reveals Trade-offs Among Oxygen Delivery, Shear, and Washout in Pediatric Roller-Pump Cardiopulmonary Bypass
Yunpeng Tu, Yi‐Ting YehiD, Alex De Lecea, Kevin A. CharetteiD, Lyubomyr BohutaiD, Vishal NigamiD, Juan C. del ÁlamoiD, Christina L. GreeneiD
Abstract
Objective: To determine how roller speed, tubing occlusion, and hematocrit jointly regulate oxygen delivery, shear, and washout in pediatric roller-pump cardiopulmonary bypass. Methods: A three-dimensional, closed-loop fluid-structure interaction model resolved roller-tube contact, tubing deformation, and blood flow. Forty simulations combined 4 roller speeds (38-95 revolutions per minute), 2 occlusion levels (77% and 83%), and 5 hematocrit levels (15%-50%). Cycle-averaged flow, oxygen delivery, peak shear stress, and viscosity-normalized relative residence time were quantified. Results: Roller compression produced retrograde leakage through the occluded gap, localized peak shear stress exceeding 100 pascals, and adjacent stagnation and recirculation. Flow increased nearly linearly with roller speed and was higher at greater occlusion, with little hematocrit dependence. Higher speed and occlusion increased peak shear stress while reducing relative residence time; higher hematocrit further increased peak shear stress. Similar oxygen delivery could arise from substantially different flow environments: among 10 conditions providing oxygen delivery of approximately 9-12 milliliters per minute, peak shear stress ranged from approximately 25 to 70 pascals. Within conditions satisfying an illustrative minimum oxygen-delivery threshold, lower hematocrit permitted speed and occlusion settings yielding lower peak shear stress and relative residence time. Conclusions: Pediatric roller-pump settings that provide similar oxygen delivery can expose blood to markedly different shear and washout environments. Joint evaluation of these metrics may guide selection of speed, occlusion, and hematocrit to meet a target oxygen delivery while reducing shear and improving washout. Keywords: Cardiopulmonary bypass; roller pump; pediatric cardiac surgery; fluid-structure interaction; shear stress; relative residence time
- Citations
- 0
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Applications of a Novel Reciprocating Positive Displacement Pump in the Simulation of Pulsatile Arterial Blood Flow
A. Menkara, A. Faryami, … C. Harris·bioRxiv·Jun 20, 2022
- Assessing the Impact of Valve Sizing and Orientation on Bioprosthetic Pulmonary Valve Hemodynamics Using In Vitro 4D-Flow MRI
Nicole Schiavone, Priya J. Nair, … Alison Lesley Marsden·Cardiovascular Engineering and TechnologyPeer-reviewed·Dec 4, 2024
- Dissecting contributions of pulmonary arterial remodeling to right ventricular afterload in pulmonary hypertension
Sunder Neelakantan, Emilio A. Mendiola, … Reza Avazmohammadi·Bioengineering & Translational MedicinePeer-reviewed·Jun 26, 2025
- A Rigid Parallel-Plate Artificial Placenta Oxygenator with a Hemocompatible Blood Flow Path
D. G. Blauvelt, N. C. Higgins, … S. Roy·bioRxiv·Aug 23, 2022
- Optimized Biomechanical Design of a Pulsatile Fontan Conduit for Congenital Heart Palliation
Nir Emuna, Zinan Hu, … Jay D. Humphrey·npj Regenerative MedicinePeer-reviewed·Jan 15, 2026
- Thoracoabdominal pressure transmission during prone and supine cardiopulmonary resuscitation in fresh-frozen human cadavers
Zachary Smith, Robert M. Lust, … Linda P. Bolin·bioRxiv·Sep 29, 2026