Two-photon targeted, quad whole-cell patch clamping robot

G. I. V. Gonzalez, P. O. Kgwarae, S. R. Schultz

PDFDOI

Abstract

We present an automated quad-channel patch-clamp technology platform for ex vivo brain slice electrophysiology, capable of both blind and two-photon targeted robotically automated patching. The robot scales up the patch-clamp singlecell recording technique to four simultaneous channels, with seal success rates for two-photon targeted and blind modes of 54% and 68% respectively. In 50% of targeted trials (where specific cells were required), at least 2 simultaneous recordings were obtained. For blind mode, most trials yielded dual or triple recordings. This robot, a milestone on the path to a true in vivo robotic multi-patching technology platform, will allow numerous studies into the function and connectivity patterns of both primary and secondary cell types.

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISJouhanneau '19Koós '21Jehasse '24Yip '24Dacre '23Sharma '21
66% semantic matchNeuroscience
Jean-Sébastien Jouhanneau, James F.A. Poulet·Frontiers in Synaptic Neuroscience·May 16, 2019· 41 citations