Multidimensional T1-T2 Relaxation Imaging In Vivo on a Portable 0.064~T MRI Scanner
Ella WilczynskiiD, Kulam Najmudeen MagdoomiD, Alexandru AvramiD, Nathan H. WilliamsoniD, Dan BenjaminiiD, Silvina G. HorovitziD, Peter J. BasseriD
Abstract
Purpose: To demonstrate the feasibility of whole-brain multidimensional T1-T2 relaxation imaging on a portable 0.064 T MRI scanner. Methods: A two-dimensional inversion-recovery, fast spin-echo acquisition was used to jointly encode T1 and T2 relaxation. Joint relaxation distributions were reconstructed voxel-wise using marginal-distribution constrained optimization (MADCO). The approach was evaluated in a quantitative relaxation MRI phantom and subsequently applied in vivo in a healthy volunteer. Joint distributions, marginal distributions, distribution-derived measures, and statistical dependencies between the two relaxation dimensions were examined. Results: Phantom joint relaxation estimates showed good overall agreement with mono-exponential relaxation measurements and previously reported values at 0.064 T. The in vivo acquisition provided whole-brain coverage over a broad range of relaxation weightings and tissue contrasts. Reconstructed joint T1-T2 distributions showed spatially organized features across the relaxation space, including broad and overlapping relaxation components that were not fully represented by single-value relaxation maps or by either 1D marginal distribution alone. Selected regions of the joint relaxation space produced component maps with distinct spatial patterns. Statistical analysis further showed a dependence between the two relaxation dimensions, supporting the presence of added information in the joint distribution that is lost when treating T1 and T2 separately. Conclusion: Whole-brain multidimensional T1-T2 relaxation imaging is clinically feasible on a portable 0.064 T MRI scanner. Joint relaxation distributions provide information beyond single-value relaxation mapping and may support further developments of quantitative multidimensional imaging at ultra-low field.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 0
- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Mesoscopic whole-brain T2*-weighted and associated quantitative MRI in healthy humans at 10.5 T
Jiaen Liu, Peter van Gelderen, … Xiaoping Wu·Magnetic Resonance in MedicinePeer-reviewed·Apr 7, 2026
- Ultra-low-field brain MRI morphometry: test-retest reliability and correspondence to high-field MRI
František Váša, Carly Bennallick, … Steven Williams·Imaging NeurosciencePeer-reviewed·Jan 1, 2025
- Time-efficient relaxation measurements by 31P MR fingerprinting in human brain at 7T
S.-I. Lim, M. S. Widmaier, … L. Xin·bioRxiv·May 23, 2022
- Comparison of test-retest reproducibility of DESPOT and 3D-QALAS for water T1 and T2 mapping
Gizeaddis Lamesgin Simegn, Borjan A. Gagoski, … Richard A.E. Edden·bioRxiv·Aug 15, 2024
- In vivo disentanglement of diffusion frequency-dependence, tensor shape, and relaxation using multidimensional MRI
Jessica T. E. Johnson, Mustafa Okan Irfanoglu, … Dan Benjamini·Human Brain MappingPeer-reviewed·May 1, 2024
- A Multicenter Preclinical MRI Study: Definition of Rat Brain Relaxometry Reference Maps
Tristan Deruelle, Frank Kober, … Michel Dojat·Frontiers in NeuroinformaticsPeer-reviewed·May 19, 2020