Proceedings of the National Academy of SciencesBiophysicsPeer-reviewed journal article
Identifying Intermolecular Interactions in Single-Molecule Localization Microscopy
Xingchi YaniD, Polly Y. YuiD, Arvind SrinivasaniD, Sohaib Abdul RehmaniD, Surabhi Kottigegollahalli SreenivasiD, Jeremy B. ConwayiD, Maxim B. PrigozhiniD
Abstract
Intermolecular interactions underlie all cellular functions, yet visualizing these interactions at the single-molecule level remains challenging. Single-molecule localization microscopy (SMLM) offers a potential solution. Given a nanoscale map of two putative interaction partners, it should be possible to assign molecules either to the class of coupled pairs or to the class of non-coupled bystanders. Here, we developed a probabilistic algorithm that allows accurate determination of both the absolute number and the proportion of molecules that form coupled pairs. The algorithm calculates interaction probabilities for all possible pairs of localized molecules, selects the most likely interaction set, and corrects for any spurious colocalizations. Benchmarking this approach across a set of simulated molecular localization maps with varying densities (up to [~] 50 molecules {micro}m-2) and localization precisions (5 to 50 nm) showed typical errors in the identification of correct pairs of only a few percent. At molecular densities of [~] 5-10 molecules {micro}m-2 and localization precisions of 20-30 nm, which are typical parameters for SMLM imaging, the recall was [~] 90%. The algorithm was effective at differentiating between non-interacting and coupled molecules both in simulations and experiments. Finally, it correctly inferred the number of coupled pairs over time in a simulated reaction-diffusion system, enabling determination of the underlying rate constants. The proposed approach promises to enable direct visualization and quantification of intermolecular interactions using SMLM.
- assay: fluorescence / live-cell microscopy
- research focus: cell-cell communication
- approach: statistical model
- Citations
- 5
- FWCI
- 1.62
- Field percentile
- top 18%
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Correlation Single Molecule Localization Microscopy (corrSMLM) Detects Fortunate Molecules for Better Localization Precision and High Signal-to-Noise Ratio
A. S, F. C. Zanacchi, P. P. Mondal·bioRxiv·Dec 29, 2022
- A method to estimate the effective point spread function of static single molecule localization microscopy images
T. R. Shaw, F. J. Fazekas, … S. L. Veatch·bioRxiv·Mar 5, 2022
- Identification and calibration of ultrabright localizations to eliminate quantification error in SMLM
- Relative enrichment - a density-based colocalization measure for single-molecule localization microscopy
A. L. Ejdrup, M. D. Lycas, … U. Gether·bioRxiv·Feb 8, 2022
- A Method for Quantifying Molecular Interactions Using Stochastic Modelling and Super-Resolution Microscopy
Keria Bermudez‐Hernandez, Sarah Keegan, … David Fenyö·Scientific ReportsPeer-reviewed·Oct 26, 2017
- Robust and fast multicolor Single Molecule Localization Microscopy using spectral separation and demixing
K. Friedl, A. Mau, … C. Leterrier·bioRxiv·Jan 23, 2023