Titanium and platinum coatings reduce inflammation induced by gold on subretinal prosthesis
Mohajeet B. BhuckoryiD, Vladimir MamchikiD, Nicharee MonkongpitukkuliD, Davis Pham-HowardiD, Viktoryia I. ShautsovaiD, Linh Mai Vu, Ludwig Galambos, Emma ButtiD, Keith MathiesoniD, Theodore I KaminsiD, Daniel PalankeriD
Abstract
Subretinal photovoltaic implants provide central vision to patients blinded by atrophic age-related macular degeneration, with acuity limited by their 100-{micro}m pixels. Higher resolution requires smaller pixels incorporating three-dimensional electrodes, which can be fabricated by gold electroplating. However, the retinal response to exposed gold remains poorly characterized. Here, we evaluated gold biocompatibility on subretinal implants in Royal College of Surgeons rats and compared it with platinum- and titanium-coated surfaces. Although in-vivo optical coherence tomography revealed no overt structural disruption, gold implants induced cellular-scale anomalies, including abnormal morphology of rod bipolar cells, microglial accumulation near the implant, and increased cell death within days after implantation. These effects occurred across flat, pillar, and honeycomb geometries, indicating a material-rather than geometry-dependent response. By contrast, platinum- and titanium-coated implants showed substantially lower loss and morphological disruption of rod bipolar cells, together with markedly reduced microglial activation. These findings indicate that exposed gold surfaces can induce acute retinal inflammation and neuronal loss, whereas conformal platinum or titanium coatings substantially improve biocompatibility. Such coatings enable the development of three-dimensional subretinal prostheses with smaller pixels for improved visual resolution.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 0
- Organism
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Electrodeposited Biocompatible Coatings for 3D Electrodes on Retinal Prostheses
Viktoryia I. Shautsova, Emma Butt, … Daniel Palanker·bioRxiv·Sep 14, 2026
- Cellular integration with a subretinal honeycomb-shaped prosthesis
Mohajeet B. Bhuckory, Zhijie Chen, … Daniel Palanker·bioRxiv·Dec 15, 2022
- 3D electronic implants in subretinal space: long-term follow-up in rodents
Mohajeet B. Bhuckory, Bing-Yi Wang, … Daniel Palanker·BiomaterialsPeer-reviewed·Dec 1, 2024
- Enhancing Prosthetic Vision by Upgrade of a Subretinal Photovoltaic Implant in situ
Mohajeet B. Bhuckory, Nicharee Monkongpitukkul, … Daniel Palanker·Nature CommunicationsPeer-reviewed·Mar 21, 2025
- Maximizing the fidelity of a photovoltaic subretinal prosthesis for human patients
Nathan Jensen, Keith Ly, … Daniel Palanker·Journal of Neural EngineeringPeer-reviewed·May 13, 2026
- Three-dimensional electro-neural interfaces electroplated on subretinal prostheses
Emma Butt, Bing-Yi Wang, … Keith Mathieson·Journal of Neural EngineeringPeer-reviewed·Feb 1, 2024