bioRxivBioengineeringPreprint
FE-DOE: Finite element informed design of experiments to optimise bioinspired melt electrowritten (MEW) polymeric heart valves
Celia HughesiD, Robert D. Johnston, Desmond McCarthy, Ewa Klusak, Emily A. Growney KalafiD, Evelyn Campbell, Caitríona LallyiD
- Citations
- 0
Abstract
Aortic stenosis is predominantly treated through transcatheter bioprosthetic heart valve implantation. The materials used in these devices, however, suffer from premature failure. Polymer heart valves have the potential to improve current commercial devices, offering materials with extended durability and customisation through fibre-reinforcement. Due to the wide range of available materials and structures, there is a need for a methodical approach to the design and optimisation of novel, bioinspired polymeric leaflets. This work presents a framework using FE and DOE tools to enable the creation of optimised bioinspired, 3D-printed, fibre-reinforced polymer leaflets using MEW. Here, FE models are created to represent MEW fibre-reinforced polymer leaflets for application in a transcatheter aortic heart valve. The behaviour of this valve under physiological loading conditions is modelled to predict valve performance and leaflet material response. These models were first used to investigate the impact of fibre orientation on valve performance and leaflet response, showing the benefit of using a bioinspired fibre reinforcement structure. Using DOE, the structural combination of MEW fibre-reinforcement and elastomeric matrix was optimised based on valve performance and leaflet stress and strain. Overall, the framework offers an efficient and versatile methodology for optimising fibre-reinforced polymer leaflets by utilising an in-silico approach to remove the need for manufacturing and testing of these devices.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Hydrogel-Polyurethane Fiber Composites With Enhanced Microarchitectural Control For Heart Valve Replacement
A. Robinson, A. Nkansah, … E. Cosgriff-Hernandez·bioRxiv·Sep 29, 2023
- Functional Performance of Electrospun Tissue-Engineered Heart Valve Scaffolds Across Geometries Under Aortic Conditions
Hannes Dillinger, Aref Saberi, … Carlijn V. C. Bouten·bioRxiv·Oct 14, 2026
- Biomimetic biaxial mechanical properties enhance hemodynamic performance and prevent adverse tissue remodelling in tissue-engineered heart valves
Bahram Mirani, Sean O. Mathew, … Craig A. Simmons·bioRxiv·Mar 2, 2025
- Effect of biomaterial stiffness on cardiac mechanics in a biventricular infarcted rat heart model with microstructural representation of in situ intramyocardial injectate
Y. D. Motchon, K. L. Sack, … T. Franz·bioRxiv·May 23, 2022
- Multi-objective optimisation of material properties and strut geometry for poly(L-lactic acid) coronary stents using response surface methodology
- Influence of Supraphysiologic Biomaterial Stiffness on Ventricular Mechanics and Myocardial Infarct Reinforcement
R. Ghanta, Y. Zhao, … K. J. Lampe·bioRxiv·May 20, 2020