Detection of the 40-Hz Auditory Steady-state Response with Optically Pumped Magnetometers
Kyung‐min AniD, Jeong Hyun ShimiD, Hyukchan Kwon, Yong‐Ho LeeiD, Kwon-Kyu YuiD, Moonyoung KwoniD, Woo Young ChuniD, Tetsu HirosawaiD, Chiaki HasegawaiD, Sumie IwasakiiD, Mitsuru KikuchiiD, Kiwoong KimiD
Abstract
Magnetoencephalography (MEG) is a functional neuroimaging technique that noninvasively detects the brain magnetic field from neuronal activations. Conventional MEG measures brain signals using superconducting quantum interference devices (SQUIDs). SQUID-MEG requires a cryogenic environment involving a bulky non-magnetic dewar and the consumption of liquid helium, which restricts the variability of the sensor array and the gap between the cortical sources and sensors. Recently, miniature optically pumped magnetometers (OPMs) have been developed and commercialized. OPMs do not require cryogenic cooling and can be placed within millimeters from the scalp. In the present study, we arranged six OPM sensors on the temporal area to detect auditory-related brain responses in a two-layer magnetically shielded room. We presented the auditory stimuli of 1-kHz pure-tone bursts with 200-ms duration and obtained the M50 and M100 components of auditory evoked fields. We delivered the periodic stimuli with a 40-Hz repetition rate and observed the gamma-band power changes and inter-trial phase coherence of auditory steady-state responses at 40 Hz. We found that the OPM sensors have a performance comparable to that of conventional SQUID-MEG sensors, and our results suggest the feasibility of using OPM sensors for functional neuroimaging and brain-computer interface applications.
02 /Specification & Classification
Click any tag to browse matching papers- Citations
- 19
- FWCI
- 1.63
- Field percentile
- top 16%
- Organism
- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A Novel, Robust, and Portable Platform for Magnetoencephalography using Optically Pumped Magnetometers
Holly Schofield, Ryan M. Hill, … Matthew J. Brookes·Imaging NeurosciencePeer-reviewed·Jan 1, 2024
- Measuring MEG closer to the brain: Performance of on-scalp sensor arrays
Joonas Iivanainen, Matti Stenroos, Lauri T. Parkkonen·NeuroImagePeer-reviewed·Dec 19, 2016
- Measurement Equivalence of On-Scalp OPM-MEG and Cryogenic MEG for Auditory and Somatosensory Cortical Mapping Across Development
William C. Gaetz, George C. O’Neill, … Orang Alem·bioRxiv·Jul 20, 2026
- The Mismatch Negativity compared: EEG, SQUID-MEG and novel 4Helium-OPMs
Tjerk P. Gutteling, Jérémie Mattout, … Françoise Lecaignard·Human Brain MappingPeer-reviewed·Sep 27, 2025
- On-scalp MEG system utilizing an actively shielded array of optically-pumped magnetometers
Joonas Iivanainen, Rasmus Zetter, … Lauri T. Parkkonen·NeuroImagePeer-reviewed·Mar 15, 2019
- Signal-to-noise ratio of event-related fields in on-scalp and off-scalp MEG
Mainak Jas, Teppei Matsubara, … Seppo P. Ahlfors·bioRxiv·Aug 21, 2026