A microfluidic platform with integrated porous membrane cell-substrate impedance spectroscopy (PM-ECIS) for biological barrier assessment
Abstract
Traditionally, biological barriers are assessed in vitro by measuring trans-endothelial/epithelial electrical resistance (TEER) across a monolayer using handheld chopstick electrodes. Implementation of TEER into organ-on-chip (OOC) setups is a challenge however, due to non-uniform current distribution and interference from biomaterials typically found in such systems. In this work, we address the pitfalls of standard TEER measurement through the application of porous membrane electrical cell-substrate impedance sensing (PM-ECIS) to an OOC setup. Gold leaf electrodes (working electrode diameters = 250, 500, 750 {micro}m) were incorporated onto porous membranes and combined with biocompatible tape to assemble microfluidic devices. PM-ECIS resistance at 4 kHz was not influenced by presence of collagen hydrogel in bottom channels, compared to TEER measurements in same devices, which showed a difference of 1723 {+/-} 381.8 {Omega} (p=0.006) between control and hydrogel conditions. A proof of concept, multi-day co-culture model of the blood-brain barrier was also demonstrated in these devices. PM-ECIS measurements were robust to fluid shear (5 dyn/cm2) in cell-free devices, yet were highly sensitive to flow-induced changes in an endothelial barrier model. Initiation of perfusion (0.06 dyn/cm2) in HUVEC-seeded devices corresponded to significant decreases in impedance at 40 kHz (p<0.01 for 750 and 500 {micro}m electrodes) and resistance at 4 kHz (p<0.05 for all electrode sizes) relative to static control cultures, with minimum values reached at 6.5 to 9.5 hours after induction of flow. Our microfluidic PM-ECIS platform enables sensitive, non-invasive, real-time measurements of barrier function in setups integrating critical OOC features like 3D co-culture, biomaterials and shear stress.
02 /Specification & Classification
Click any tag to browse matching papers- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Porous membrane electrical cell-substrate impedance spectroscopy for versatile assessment of biological barriers in vitro
O. Chebotarev, A. Ugodnikov, C. A. Simmons·bioRxiv·Jun 26, 2023
- Sensitivity and validation of porous membrane electrical cell substrate impedance spectroscopy (PM-ECIS) for measuring endothelial barrier properties
A. Ugodnikov, O. Chebotarev, … C. A. Simmons·bioRxiv·Jul 11, 2023
- A plug-and-play transepithelial/transendothelial electric resistance (TEER)-upgraded organ-on-chip system to measure barrier dynamics in real-time
Tim Kaden, Sophie Besser, … Sándor Nietzsche·bioRxiv·Apr 10, 2025
- Modular Integration of Impedance Sensing for Real-Time Assessment of Barrier Integrity
Sami Farajollahi, Mehran Mansouri, … Vinay V. Abhyankar·bioRxiv·Mar 10, 2026
- Sub-second Extracellular Impedance Measurement of Epithelial Cell Monolayers using Step Excitations and Time-domain Analysis
Rongming Guo, Athena J. Chien, … Craig R. Forest·Scientific ReportsPeer-reviewed·May 22, 2026
- Method for Extracellular Electrochemical Impedance Spectroscopy on Epithelia
Athena J. Chien, Colby F. Lewallen, … Craig R. Forest·bioRxiv·Apr 2, 2025