henhaobio
Sign in

Biomedical Optics ExpressBiophysicsPeer-reviewed journal article

Mechanical de-skewing enables high-resolution imaging of thin tissue slices with a mesoSPIM light-sheet microscope

Steven Milan Moreno, Sharika MohananiD, Ahmed Elnageh, Erin Boland, Lewis Williamson, Camilla OliantiiD, Leonardo SacconiiD, Godfrey L. SmithiD, Eline HuethorstiD, Marie Caroline MüllenbroichiD

PDFDOIPreprint

Abstract

Optical clearing combined with light-sheet microscopy enables high-resolution imaging of extended tissue at scale. However, standard mesoSPIM systems are optimised for intact organs and are not suited to thin tissue slices. We present a no-shear imaging protocol using obliquely mounted samples held between refractive-index-matched slides in a 3D-printed frame. This enables mechanical de-skewing during acquisition, minimising post-processing requirements. We demonstrate feasibility in fluorescent bead phantoms and rabbit heart tissue, achieving a 4.8x reduction in processing time and a 1.5x improvement in axial resolution compared to the shear-method. The no-shear method extends mesoSPIMs utility to fragile, laterally extended tissue sections.

Citations
2
FWCI
0.76
Field percentile
top 28%

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.