bioRxivSystems BiologyPreprint
Cheap, Robust and Versatile Two-Dimensional Chromatography system for Proteomics of Nanogram Scale Samples
Eduardo S. KitanoiD, Gareth A. A. NisbetiD, Yana DemyanenkoiD, Katarzyna M. KowalczykiD, Louisa IseliniD, Stephen Cross, Alfredo CastellóiD, Shabaz MohammediD
Abstract
In this work we describe a low loss fractionation system comprised of a reconfigured Evosep One LC system for the first dimension and a repurposed 3D-printer as a fraction collector. The setup operates as a high-pH fractionation system capable of effectively working with nanogram scales of lysate digests. The 2D RP-RP system demonstrated superior proteome coverage over single-shot data-dependent acquisition (DDA) analysis using only 5 ng of human cell lysate digest with performance increasing with increasing amounts of material. We found that the fractionation system allowed over 70% signal recovery at the peptide level and, more importantly, we observed over 30% increase on protein level intensity which indicates the complexity reduction afforded by the system outweighs the sample losses endured. The application of data-independent acquisition (DIA) and wide window acquisition (WWA) to fractionated samples allowed more than 8,000 proteins to be identified from 50 ng of material. The utility of the 2D system was further investigated for phosphoproteomics (>21,000 phosphosites from 50 g starting material) and pull-down type experiments and showed substantial improvements over single-shot experiments. We show that the 2D RP-RP system is highly versatile and powerful tool for many proteomics workflows.
- Citations
- 2
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A novel LC system embeds analytes in pre-formed gradients for rapid, ultra-robust proteomics
N. Bache, P. E. Geyer, … M. Mann·bioRxiv·May 15, 2018
- A high-sensitivity low-nanoflow LC-MS configuration for high-throughput sample-limited proteomics
R. Zheng, M. Matzinger, … K. Mechtler·bioRxiv·Apr 27, 2023
- High-performance proteomics at any chromatographic flow rate
Giorgi Tsiklauri, Runsheng Zheng, … Bernhard Küster·bioRxiv·Apr 10, 2025
- High-flow Nano-chromatography Columns Facilitate Rapid and High-quality Proteome Analysis on Standard Nano-LC Hardware
C. S. Hughes, H. Adomat, … G. B. Morin·bioRxiv·Dec 19, 2018
- Data-Dependent Acquisition with Precursor Coisolation Improves Proteome Coverage and Measurement Throughput for Label-Free Single-Cell Proteomics
T. Truong, S. M. Johnston, … R. T. Kelly·bioRxiv·Oct 18, 2022
- Comprehensive online two-dimensional nanoLCxCZE-MS for deep top-down proteomics
Tobias Waldmann, Philipp T. Kaulich, … Christian Neusüß·bioRxiv·May 18, 2026