Highly scalable and standardized organ-on-chip platform with TEER for biological barrier modeling
H. T. H. Nguyen, S.-L. Rissanen, M. Peltokangas, T. Laakkonen, J. Kettunen, L. Barthod, R. Sivakumar, A. Palojarvi, P. Junttila, J. Talvitie, M. Bassis, S. L. Nickels, J. C. Schwamborn, S. Mosser, P. Singh
Abstract
The development of new therapies is hampered by the lack of predictive, and patient-relevant in vitro models. Organ-on-chip (OOC) technologies can potentially recreate physiological features and hold great promise for tissue and disease modeling. However, the non-standardized design of these chips and perfusion control systems has been a barrier to quantitative high-throughput screening (HTS). Here we present a scalable OOC microfluidic platform for applied kinetic in vitro assays (AKITA) that is applicable for high, medium, and low throughput. Its standard 96-well plate and 384-well plate layouts ensure compatibility with existing laboratory workflows and high-throughput data collection and analysis tools. The AKITA plate is optimized for the modeling of vascularized biological barriers, primarily the blood-brain barrier, skin, and lung, with precise flow control on a custom rocker. The integration of trans-epithelial electrical resistance (TEER) sensors allows rapid and repeated monitoring of barrier integrity over long time periods. Together with automated confocal imaging and compound permeability testing analyses, we demonstrate the flexibility of the AKITA platform for establishing human-relevant models for preclinical drug and precision medicines efficacy, toxicity, and permeability under physiological conditions.
02 /Specification & Classification
Click any tag to browse matching papersReproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- A plug-and-play transepithelial/transendothelial electric resistance (TEER)-upgraded organ-on-chip system to measure barrier dynamics in real-time
Tim Kaden, Sophie Besser, … Sándor Nietzsche·bioRxiv·Apr 10, 2025
- A microfluidic platform with integrated porous membrane cell-substrate impedance spectroscopy (PM-ECIS) for biological barrier assessment
Alisa Ugodnikov, Joy Lu, … Craig A. Simmons·bioRxiv·Nov 25, 2023
- Endothelium-on-a-chip for permeability studies in diabetic and diabetic kidney disease milieu
Vika Telle, Pim de Haan, … Elisabeth M. J. Verpoorte·bioRxiv·Sep 22, 2026
- Engeneering the neurovascular unit: a novel sensorized microfluidic platform to study barrier function and maturation
Ludovica Montesi, Davide Lattanzi, … Rossana Rauti·bioRxiv·Oct 30, 2025
- Democratizing Organ-on-Chip Technologies with a Modular, Reusable, and Perfusion-Ready Microphysiological System
Daniel J. Minahan, Katherine M. Nelson, … Jason P. Gleghorn·Advanced Healthcare MaterialsPeer-reviewed·Sep 10, 2025
- Pump-less, recirculating organ-on-a-chip (rOoC) platform
M. Busek, A. Aizenshtadt, … S. Krauss·Lab on a ChipPeer-reviewed·Jan 1, 2023