Utah Array Characterization and Histological Analysis of a Multi-Year Implant in Non-Human Primate Motor and Sensory Cortices
P. R. Patel, E. J. Welle, J. G. Letner, H. Shen, A. J. Bullard, C. M. Caldwell, J. M. Richie, H. E. Thayer, P. G. Patil, D. Cai, C. A. Chestek
Abstract
The Utah array is widely used in both clinical studies and neuroscience. It has a strong track record of safety. However, it is also known that implanted electrodes promote the formation of scar tissue in the immediate vicinity of the electrodes, which negatively impacts the ability to record neural waveforms. This scarring response has been primarily studied in rats and mice, which may have a very different response than primate brain. Here, we present a rare nonhuman primate histological dataset (n=1 rhesus macaque) obtained 848 and 590 days after implantation in two brain hemispheres. For 2 of 4 arrays that remained within the cortex, NeuN was used to stain for neuron somata at 3 different electrode depths. Images were filtered and denoised, with neurons then counted in the vicinity of the arrays as well as a nearby section of control tissue. Additionally, 3 of 4 arrays were imaged with a scanning electrode microscope (SEM) to evaluate any materials damage that might be present. Overall, we found a 63% percent reduction in the number of neurons surrounding the electrode compared to control areas. In terms of materials, the arrays remained largely intact with metal and Parylene C present, though tip breakage and cracks were observed on many electrodes. Overall, these results suggest that the tissue response in the nonhuman primate brain shows similar neuron loss to previous studies using rodents. Electrode improvements, for example using smaller or softer probes, may therefore substantially increase the neuronal recording yield in primate cortex.
02 /Specification & Classification
Click any tag to browse matching papers- Organism
- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Repeatable, low-drift recordings in behaving non-human primates using flexible microelectrodes
Daniel P. Woods, Grace Adams, … Daniel L. Gonzales·Nature CommunicationsPeer-reviewed·Sep 9, 2026
- Structural and functional changes of pyramidal neurons at the site of an implanted microelectrode array in rat primary motor cortex
B. A. Gregory, C. H. Thompson, … E. K. Purcell·bioRxiv·Sep 15, 2022
- Material Damage to Multielectrode Arrays after Electrolytic Lesioning is in the Noise
Alice Tor, Stephen E Clarke, … Paul Nuyujukian·bioRxiv·Mar 26, 2025
- Cortical Response to Acute Implantation of the Utah Optrode Array in Macaque Cortex
Adrián Villamarin‐Ortiz, Christopher F. Reiche, … Alessandra Angelucci·Advanced Healthcare MaterialsPeer-reviewed·Sep 15, 2025
- Large-scale high-density brain-wide neural recording in nonhuman primates
Eric M. Trautmann, Janis Karan Hesse, … T.D. Harris·Nature NeurosciencePeer-reviewed·Jun 23, 2025
- Intracortical probe arrays with silicon backbone and microelectrodes on thin polyimide wings enable long-term stable recordings in vivo
Antje Kilias, Yu‐Tao Lee, … Patrick Ruther·Journal of Neural EngineeringPeer-reviewed·Nov 15, 2021