Sign in
BiophysicsbioRxiv
Preprint

Alternating Angle Milling Suppresses Streaking Artifacts in FIB-SEM Imaging

Gleb E. ShtengeliD, Wei QiuiD, Jesse AaroniD, Arthur S. Crowe, Alexey A. PoliloviD, Katerina KarkaliiD, Christopher K. E. BleckiD, Harald F. HessiD

Abstract

Focused ion beam scanning electron microscopy (FIB-SEM) (1-3) has been used in life sciences to produce large volumetric datasets with high resolution information on ultrastructure of biological organisms. 3D image acquisition is accomplished by serial removal of thin layers of material using focused ion beam (FIB) milling followed by scanning electron microscopy (SEM) imaging. One of the challenges in the standard FIB-SEM imaging protocol is that FIB milling results in characteristic artifacts, known as "streaks" or "curtains". These streaks are caused by non-uniform material removal forming long straight trenches parallel to the FIB milling direction. These artifacts get worse along the milling direction and ultimately limit size of the SEM field of view. Various methods have been proposed to mitigate the streaks in acquired images. While these techniques often provide noticeable visual improvement, the underlying problem remains. The structural information in the "streaked" areas is lost due to non-uniform material removal during milling and cannot be fully recovered. We propose a simple modification allowing for a significant reduction of milling non-uniformities of streaks. We demonstrate the effectiveness of this approach on various samples.

PDFDOI

02 /Specification & Classification

Click any tag to browse matching papers

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISLi '25Rottmayer '24Brenner '25Gorelick '25Hall '26Velazco '25
72% semantic matchBioinformatics
Yicong Li, Yuri Kreinin, Siyu Huang+4 more·bioRxiv·Sep 29, 2025