henhaobio
Sign in

Abstract

We describe the design and implementation of a drop on fixed target method for time-resolved serial crystallography at both synchrotron and XFEL facilities. A piezoelectric droplet dispensing pipette is employed for addition of picolitre volume (40 - 90 pL) aqueous droplets, containing (co-)substrate(s) or ligand(s), onto enzyme microcrystals immobilised on a solid support. The system was tested with various enzyme systems, including lysozyme and two -lactamases, CTX-M-15 and AmpCEC. Mitigation strategies for cross-well contamination, including the implementation of interleaved controls, are described; the overall performance of the system at synchrotron and X-ray free electron laser facilities was evaluated. This drop on fixed target method is a reliable framework for time-resolved crystallography and will improve the consistency of measurements across facilities.

Citations
1
FWCI
1.36
Field percentile
top 28%

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.

Sign in to file a reproducibility report.

Similarity Map
SelectedSame fieldCross-field
THISStubbs '24Aller '25Kang '26Bosman '26Ghosh '26Gonen '23
77% semantic matchBiochemistry
J. Stubbs, T. Hornsey, N. Hanrahan+11 more·bioRxiv·Jan 12, 2024