Fitting a square beam in a square camera: novel condensor apertures for low-dose transmission electron microscopy
Abstract
In transmission electron microscopy (TEM) cameras are square or rectangular but beams are round. With a beam size chosen to fill the camera at a given image magnification, the circular lobes of the beam will extend beyond the cameras field of view and irradiate areas that are not acquired on the camera, damaging and precluding them from future acquisitions if the sample is beam sensitive. In this paper we present development of condenser aperture plates for TEM that have square and rectangular apertures which improve the efficiency of sample area usage by 44% or greater in low dose TEM applications. We demonstrate that the use of these apertures is compatible with high-resolution cryogenic (cryo) TEM by reconstructing sub 2 [A] apo-ferritin models from a datasets recorded with both square and rectangular apertures. Moreover the design of our aperture plates should improve the flexibility of 2 condenser systems for cryo-TEM acquisitions with multiple shots per hole by tailored matching of beam sizes to camera sizes at each magnification.
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Square beams for optimal tiling in TEM
E. Y. D. Chua, L. M. Alink, … A. de Marco·bioRxiv·Jul 29, 2023
- Selecting optimal support grids for super-resolution cryogenic correlated light and electron microscopy
Mart G.F. Last, Maarten W. Tuijtel, … Thomas Harry Sharp·Scientific ReportsPeer-reviewed·May 22, 2023
- Imaging large fields-of-view at high resolution in cryo-ET with square beam montaging
Eugene Y. D. Chua, Hamidreza Rahmani, … Alex de Marco·bioRxiv·Aug 28, 2026
- Radiation dose effects in correlative X-ray / cryo-electron microscopy of frozen hydrated biological samples
Thorsten B. Blum, Vincent Oliéric, … Volodymyr M. Korkhov·Acta Crystallographica Section D Structural BiologyPeer-reviewed·Feb 25, 2026
- Whole cryo-FIB lamella acquisition at high magnification with rectangular beam montage cryo-electron tomography
Hamish G. Brown, Sepideh Valimehr, … Eric G. Hanssen·bioRxiv·Sep 27, 2026
- A method to minimize condenser lens-induced hysteresis effects in a JEOL JEM-3200FSC microscope to enable stable cryoEM low-dose operations
M. Jason de la Cruz, Michael Martynowycz, … Tamir Gonen·bioRxiv·Jun 21, 2017