A Modified Injector and Sample Acquisition Protocol Can Improve Data Quality and Reduce Inter-Instrument Variability of the Helios Mass Cytometer
B. H. Lee, G. Kelly, S. Bradford, M. Davila, X. V. Guo, E.-a. D. Amir, E. M. Thrash, M. Solga, J. Lannigan, B. Sellers, J. Candia, J. Tsang, R. Montgomery, S. Tamaki, T. Sigdel, M. Sarwal, L. Lanier, Y. Tian, C. Kim, D. Hinz, B. Peters, A. Sette, A. Rahman
Abstract
Mass cytometry is a powerful tool for high dimensional single cell characterization. Since the introduction of the first commercial CyTOF mass cytometer by DVS Sciences in 2009, mass cytometry technology has matured and become more widely utilized, with sequential platform upgrades designed to address specific limitations and to expand the capabilities of the platform. Fluidigms 3rd-generation Helios mass cytometer introduced a number of upgrades over the previous CyTOF2. One of these new features is a modified narrow bore sample injector that generates smaller ion clouds, which is expected to improve sensitivity and throughput. However, following rigorous testing we find that the narrow-bore sample injector may have unintended negative consequences on data quality and result in lower median signal intensities and higher coefficients of variation in antibody expression. We describe an alternative Helios acquisition protocol using a wider bore injector, which largely mitigates these data quality issues. We directly compare these two protocols in a multi-site study of 10 Helios instruments across 7 institutions and show that the modified protocol improves data quality and reduces inter-instrument variability. These findings highlight and address an important source of technical variability in mass cytometry experiments that is of particular relevance in the setting of multi-center studies.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Research Focus
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- Channel crosstalk correction in suspension and imaging mass cytometry
Stéphane Chevrier, Helena Lucia Crowell, … Bernd Bodenmiller·Cell SystemsPeer-reviewed·Mar 28, 2018
- Standardization of suspension and imaging mass cytometry readouts for clinical decision making
R. Casanova, S. Xu, … S. Chevrier·bioRxiv·Mar 19, 2023
- A simple strategy for sample annotation error detection in cytometry datasets
M. Smithmyer, A. E. Wiedeman, … C. Speake·bioRxiv·Oct 26, 2021
- CytoScan: Automated detection of technical anomalies for cytometry quality control
Tim R. Mocking, Felix Zwolle, … Costa Bachas·Cytometry Part APeer-reviewed·Jul 6, 2026
- Development of a Comprehensive Antibody Staining Database using a Standardized Analytics Pipeline
E.-a. D. Amir, B. Lee, … A. H. Rahman·bioRxiv·May 2, 2019
- Agile workflow for interactive analysis of mass cytometry data
Julia Casado, Oskari Lehtonen, … Sampsa Hautaniemi·BioinformaticsPeer-reviewed·Oct 27, 2020