Sign in
MicrobiologybioRxiv

Comparison of SARS-CoV2 N gene real-time RT-PCR targets and commercially available mastermixes

J. R. Brown, D. O'Sullivan, R. P. Pereira, A. S. Whale, E. Busby, J. Huggett, K. Harris

Abstract

We aim to test four one-step RT real-time mastermix options for use in SARS-CoV2 real-time PCR, with three primer/probe assays targeting the N gene. The lower limit of detection is determined using a SARS CoV2 N gene RNA transcript dilution series (to 1 copy/{micro}l) and verified using 74 nose and throat swabs. The N2 assay demonstrates the most sensitive detection of SARS-Cov-2 RNA. Three of the four mastermixes performed well, with the Takara One Step PrimeScript III RT-PCR Kit mastermix demonstrating improved performance at the lower limit of detection.

PDFDOI

02 /Specification & Classification

Click any tag to browse matching papers

Reproducibility ledger

Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.