A highly-selective chloride microelectrode based on a mercuracarborand anion carrier
M. DiFranco, M. Quinonez, R. M. Dziedzic, A. M. Spokoyny, S. C. Cannon
Abstract
The chloride gradient plays an important role in regulating cell volume, membrane potential, pH, secretion, and the reversal potential of inhibitory GABAA receptors. Measurement of intracellular chloride activity, [Formula], using liquid membrane ion-selective microelectrodes (ISM), however, has been limited by the physiochemical properties of Cl ionophores which have caused poor stability, drift, sluggish response times, and interference from other biologically relevant anions. Most importantly, intracellular HCO3- may be up to 4 times more abundant than Cl- (e.g. skeletal muscle) which places severe constraints on the required selectivity of a Cl - sensing ISM.\n\nPreviously, a sensitive and highly-selective Cl sensor was developed in a polymeric membrane electrode (Badr et al. 1999) using a trinuclear Hg(II) complex containing carborane-based ligands, [9]-mercuracarborand-3, or MC3 for short. Here, we have adapted the use of the MC3 anion carrier in a liquid membrane ion-selective microelectrode and show the MC3-ISM has a linear Nernstian response over a wide range of aCl (0.1 mM to 100 mM), is highly selective for Cl over other biological anions or inhibitors of Cl transport, and has a response time of less than 5 sec. Importantly, over the physiological range of aCl (1 mM to 100 mM) the potentiometric response of the MC3-ISM is insensitive to HCO3- or changes in pH. Finally, we demonstrate the biological application of an MC3-ISM by measuring intracellular aCl, and the response to an external Cl-free challenge, for an isolated skeletal muscle fiber.
02 /Specification & Classification
Click any tag to browse matching papers- Assay
- Approach
Reproducibility ledger
Nobody has reported trying to reproduce this paper yet. Did you replicate an experiment or run the code? Your report saves the next lab a week.
Sign in to file a reproducibility report.
Related papers
- KCC2-dependent Steady-state Intracellular Chloride Concentration and pH in Cortical Layer 2/3 Neurons of Anesthetized and Awake Mice
Juan Carlos Boffi, Johannes Knabbe, … Thomas Kuner·Frontiers in Cellular NeurosciencePeer-reviewed·Jan 24, 2018
- Displacement of extracellular chloride by immobile anionic constituents of the brain's extracellular matrix
Kieran P. Normoyle, Kyle P. Lillis, … Kevin J. Staley·The Journal of PhysiologyPeer-reviewed·Dec 2, 2024
- Biophysical models reveal the relative importance of transporter proteins and impermeant anions in chloride homeostasis
Kira M. Düsterwald, Christopher Brian Currin, … Joseph V. Raimondo·eLifePeer-reviewed·Sep 27, 2018
- A Four Electrode Method to Study Dynamics of Ion Activity and Transport in Skeletal Muscle Fibers
J. A. Heiny, S. C. Cannon, M. C. DiFranco·bioRxiv·May 11, 2019
- Backbone amides are key determinants of Cl- selectivity in CLC ion channels
L. Leisle, K. Lam, … A. Accardi·bioRxiv·Jul 1, 2022
- Chloride ions evoke taste sensations by binding to the extracellular ligand-binding domain of sweet/umami taste receptors
N. Atsumi, K. Yasumatsu, … A. Yamashita·bioRxiv·Feb 23, 2022